Using
a
Spectrum
Analyzer
This
section
provides
the
probe
adjustment
procedure
using
a
spectrum
analyzer
.
Equipment
Required.
Spectrum
Analyzer:
Any
Compatible
Spectrum
Analyzer
50
F
eedthrough:
HP
10100C
A
dapters:
as
necessary
Procedure
.
1.
Connect
the
equipment
as
shown
in
Figure
3-7.
Note
that
the
probe
power
plug
must
be
connected
before
the
HP
41802A's
N-type
connector
.
Connect
the
ground
lead
rmly
to
provide
good
electrical
bond.
Note
If
the
spectrum
analyzer
has
a
tracking
generator
output,
use
the
tracking
generator
output
connector
instead
of
the
calibration
output
connector
.
Then
set
the
output
level
to
010
dBm.
Figure
3-7.
Probe
A
djustment
Setup
Using
Spectrum
Analyzer
2.
Setup
the
spectrum
analyzer
as
follows:
Center
Frequency:
Same
as
Calibration
Output
frequency
.
Frequency
Span:
50
kHz.
Attenuation:
10
dB
.
3.
T
ouch
the
center
conductor
of
the
feedthrough
with
the
probe
tip
.
4.
A
djust
the
trimmer
component
(through
opening
in
probe
assembly
cover)
until
the
peak
signal
level
is
as
close
to
`calibration
output
level
0
20
dB'
as
possible
.
3-10
Operation
Содержание 41802A