Operating
Check
Using
a
Network/Spectrum
Analyzer
This
section
provides
the
HP
41802A
operating
check
procedure
using
the
HP
4195A
network/spectrum
analyzer
.
Equipment
Required
Network/Spectrum
Analyzer:
HP
4195A
BNC(m)-BNC(m)
Cable
,
30
cm:
PN
8120-1838
A
dapters:
as
necessary
Procedure
1.
Connect
the
equipment
as
shown
in
Figure
3-4.
Note
that
the
probe
power
plug
must
be
connected
before
the
HP
41802A's
N-type
connector
.
Figure
3-4.
Operating
Check
Setup
Using
Network/Spectrum
Analyzer
2.
Set
up
the
Network/Spectrum
analyzer
as
follows:
Conguration:
Spectrum
Center
Frequency:
1
MHz
Frequency
Span:
0
Hz
P
ort
Select:
Source
Channel
1
on
Amplitude
(S1):
026
dBm
Ref
Attenuator
(R1):
10
dB
3.
Conrm
that
the
signal
level
is
`020
dB
6
5
dB
.'
Operation