Single Triggering
In the single trigger mode (front panel only),
you can manually trigger the function generator by pressing the
front-panel
Single
key. The function generator outputs one burst or
initiates one sweep each time you press the key. The
Trig
annunciator turns on when the function generator is waiting for
a single trigger.
The front-panel
Single
key is disabled when in remote
.
External Triggering
In this trigger mode, the function generator will
accept a hardware trigger applied to the
Ext Trig
terminal. The function
generator outputs one burst or initiates one sweep each time
Ext Trig
receives the
rising edge
of a
TTL
pulse.
See also “Ext Trig/FSK/Burst Input Terminal,” on page 102.
•
Front-Panel Operation
: The external trigger mode is like the single
trigger mode except that you apply the trigger to the
Ext Trig
terminal. Pressing
Single
to enable the single trigger mode also
enables the external trigger mode. The
Trig
annunciator turns on
when the function generator is waiting for an external trigger.
The front-panel
Single
key is disabled when in remote
.
•
Remote Interface Operation
:
TRIGger:SOURce EXTernal
Chapter 3 Features and Functions
Triggering
100
Содержание 33120A