Table 3-1. Performance Test Record (Continued)
Paragraph No.
Test
Results
3-9
Voltage Measurement Accuracy
(continued)
Upper Limit
Lower Limit
-13.7V
-14.3V
-3.43V
-3.57V
-1.37V
-1.43V
-137.0mV
-143.0mV
143.0mV
137.0mV
1.43V
1.37V
3.57V
3.43V
14.3V
13.7V
Channel 2
Zero Input
Measured
Actual
__________
__________
__________V
__________
__________
__________V
__________
__________
__________V
__________
__________
__________mV
__________
__________
__________mV
__________
__________
__________V
__________
__________
__________V
__________
__________
__________V
3-10
Offset Accuracy
Zero-input Offset-Channel 1
Offset Accuracy
DC Input Offset-Channel 1
Offset Accuracy
Zero-input Offset-Channel 2
Offset Accuracy
DC Input Offset-Channel 2
V/Div Setting Reading Actual
4V/Div 0.00V
±
320.0mV ___________V
1V/Div 0.00V
±
80.0mV ___________V
100mV/Div 0.00V
±
8.0mV ___________V
10mV/Div 0.00V
±
800.0
µ
V ___________V
Supply Min Max Actual
-35.00V -35.4V -34.6V -__________V
-10.00V -10.1V -9.90V -__________V
-2.00V -2.02V -1.98V -__________V
+2.00V +1.98V +2.02V +__________V
+10.00V +9.90V +10.1V +__________V
+35.00V +34.6V +35.4V +__________V
V/Div Setting Reading Actual
4V/Div 0.00V
±
320.0mV ___________V
1V/Div 0.00V
±
80.0mV ___________V
100mV/Div 0.00V
±
8.0mV ___________V
10mV/Div 0.00V
±
800.0
µ
V ___________V
Supply Min Max Actual
-35.00V -35.4V -34.6V -__________V
-10.00V -10.1V -9.90V -__________V
-2.00V -2.02V -1.98V -__________V
+2.00V +1.98V +2.02V +__________V
+10.00V +9.90V +10.1V +__________V
+35.00V +34.6V +35.4V +__________V
3-11
Bandwidth
≤ −
3.0dB
Channel 1:
Actual Reading= ________ dB
Channel 2:
Actual Reading = ________ dB
HP 16532A - Performance Tests
3-18