Table 1-3. Recommended Test Equipment
EQUIPMENT
CRITICAL SPECIFICATIONS
RECOMMENDED MODEL
USE*
Signal Generator
Frequency: 1 - 250 MHz at approx .
170 mV RMS resolution
Output Accuracy:
±
1 dB
1 MHz time base accuracy 0.25 ppm
HP 8656B Option 001
P
Dc Power Supply
Range:
−
35.000 to + 35.000 Vdc,
±
1 mV HP 6114A
P
Digital Multimeter
0.1 mV resolution
Accuracy: better than 0.005%
Resistance measurement: better than
0.25% accuracy
HP 3458A
P
Power Meter/Sensor
1 - 250 MHz
±
3% accuracy
HP 436/8482A
P
Power Splitter
Outputs differ by < 0.15 dB
HP 11667B
P
Blocking Capacitor
50
Ω
BNC (m-to-f), 0.18
µ
F,
±
200 V
HP 10240B
P
Adapter
50
Ω
BNC (m)(m)
HP 1250-1236
C
Adapter
50
Ω
Type N(m)-to-BNC(m)
HP 1250-0082
P
Adapter
50
Ω
BNC(f)-to-Dual Banana Plug
HP 1251-2277
P
Adapter
50
Ω
Type N(m)-to-BNC(f)
HP 1250-0780
P
Tee Adapter
50
Ω
BNC (m)(f)(f)
HP 1250-0781
P,C
Cable
Type N(m-to-m) 24-inch
HP 11500B
P
Cable
50
Ω
BNC (m-to-m) 48-inch
HP 10503A
P
Cable
50
Ω
BNC (m-to-m) 9 inch
HP 10502A
C
Cable
50
Ω
BNC-SMB
HP 16532-61601
C
* P= Performance Tests C= Calibrations
HP 16532A - General Information
1-9