PhacoJack
R
THE Phacosystem
32
I/A HANDPIECE
SYSTEM TEST (IRRIGATION FLOW)
FIGURE 24
TROUBLESHOOTING TIPS
* MAKE SURE ALL CONNECTIONS TO THE
HANDPIECE ARE SECURE.
* AVOID SEVERE BENDS / KINKS IN ALL TUBING
* MAKE SURE THE HANDPIECE IS FREE
OF OBSTRUCTIONS.
NORMAL
COLLAPSE
WARNING!
SET CONSOLE TO I/A MODE PER SURGEONS
INSTRUCTIONS. I/A VAC & RATE SETTINGS
FOR SYSTEM TEST REPRESENT THE MAXIMUM
THAT THE SURGEON WILL USE DURING
SURGERY. IF THESE SETTINGS ARE INCREASED DURING
SURGERY AND / OR THE IRRIGATION CONTAINER IS
LOWERED, THE SYSTEM TEST MUST BE REPEATED IN
ORDER TO GUARANTEE SAFE OPERATING CONDITIONS.
PROCEDURE
1 PINCH IRRIGATION TUBING WITH CLAMP (HEMOSTAT).
2 DEPRESS FOOTPEDAL TREADLE TO SECOND DETENT (DETENT 2) AS SHOWN.
3 NOTE PUMP START AND STOP MOTION. IF PUMP DOES NOT REMAIN
STOPPED OR THERE IS INTERMITTENT MOTION, STOP. DO NOT PROCEED
UNTIL PUMP REMAINS STOPPED FOR > 10 SEC.
4 NOTE TEST CHAMBER COLLAPSE. DO NOT RELEASE FOOTPEDAL TREADLE.
IF NO COLLAPSE, STOP. DO NOT PROCEED UNTIL COLLAPSE IS OBSERVED.
5 RELEASE CLAMP (HEMOSTAT)
6 NOTE REINFLATION OF TEST CHAMBER. IF THERE IS NO REINFLATION, STOP.
DO NOT PROCEED UNTIL REINFLATION IS OBSERVED.
7 RELEASE FOOTPEDAL TREADLE.
MODE
CAUTERY
I/A VAC
U/S VAC
RATE
U/S POWER
PATIENT EYE