PhacoJack
R
THE Phacosystem
41
VITRECTOR
SYSTEM SETUP (PRIME)
FIGURE 33
* SCRUB NURSE
* CIRCULATING NURSE
DRAINAGE BAG
MAXIMUM CAPACITY 500ml
ACTUATOR TUBING
VIT FITTING
PROCEDURE
24 DEPRESS FOOTPEDAL TREADLE TO FIRST DETENT (DETENT 1)
AS SHOWN (IRRIGATION ONLY) AND HOLD UNTIL
TEST CHAMBER IS FILLED ~ 3/4 FULL.
25 PLACE TEST CHAMBER ON VITRECTOR.
26 DEPRESS FOOTPEDAL TREADLE TO SECOND DETENT (DETENT 2)
AS SHOWN (IRRIGATION & ASPIRATION) AND HOLD UNTIL
ALL TUBING IS PRIMED.
27 CONNECT ACTUATOR TUBING TO CONSOLE AS SHOWN.
28. RELEASE FOOTPEDAL TREADLE.
DRAINAGE BAG
MAXIMUM CAPACITY 500ml