PhacoJack
R
THE Phacosystem
38
VITRECTOR
SYSTEM SETUP (IRRIGATION TUBING SET)
FIGURE 30
PROCEDURE
9 PLACE IRRIGATION SLEEVE ONTO VITRECTOR TIP.
10 INSERT IRRIGATION TUBING SET FITTING INTO VITRECTOR.
11 PASS DRIP CHAMBER TO CIRCULATING NURSE. DO NOT PERMIT DRIP
CHAMBER SPIKE TO TOUCH ANY NON-STERILE SURFACE WHILE HANDLING.
12 INSERT SPIKE INTO IRRIGATING SOLUTION BOTTLE. DO NOT
PERMIT DRIP CHAMBER SPIKE TO TOUCH ANY NON-STERILE
SURFACE WHILE INSERTING SPIKE INTO BOTTLE.
13 SET BOTTLE HEIGHT AS SHOWN.
14 PASS SILICONE TUBING PORTION OF IRRIGATION TUBING SET
TO CIRCULATING NURSE.
15 DEPRESS FOOTPEDAL TREADLE TO FIRST DETENT (DETENT 1)
AS SHOWN.
16 INSERT TUBING INTO PINCH VALVE.
17 RELEASE FOOTPEDAL TREADLE.
PINCH VALVE TOP VIEW
TUBING INSTALLED
* CIRCULATING NURSE
* SCRUB NURSE
65cm
(25in)
(PINCH VALVE)
VITRECTOR