Maintenance and water quality
Make sure that the medium is clean in order to avoid deposits such as lime, rust and dirt in the
measuring module. The water flowing through should be free of suspended particles and meet
the following requirements:
Min.
–
Max.
Unit
pH value
6.5
–
9.5
-
Solid substances
free
mg/l
Electrical conductivity (20°C)
200 - 2000
µS/cm
Fe / Cu
0 - 1
mg/l
Total hardness
5 - 12
°dH
Using compressed air to clean or empty the measuring module can damage the measuring
components. Cleaning brushes or similar objects should also not be used to remove any
deposits. The individual components can only be cleaned carefully with a cloth once the
measuring module has been disassembled.
We recommend that you return the device to Hotset GmbH for maintenance and servicing.