System Operating Manual
29
charged upon receipt. Do not place device in service if it fails
the self test.
CAUTION:
Do not operate the Plum XL with the battery
removed. The use of a properly maintained and charged
battery ensures proper operation. As always, in the event
of an AC power interruption or failure, verify infusion
pump settings.
Only LifeCare 5000 PlumSet administration sets can be used
with this device.
After the cassette door is closed, turn the control dial to SET
RATE. Verify screen display (see Self Test). If the Plum XL
screen does not match exactly, remove the device from service
and contact appropriate personnel.
Consult drug container labeling to confirm drug compatibility,
concentration, delivery rates, and volumes are all suitable for
secondary or piggyback delivery mode.
Setting the primary rate greater than the secondary rate will
result in a more rapid infusion of any residual secondary drug
remaining in the line and the cassette.
Backpriming is not recommended for reconstituting secondary
containers containing dry powders.
To avoid pressurization when backpriming into a syringe or a
vial, the user must ensure that these containers have sufficient
empty space to accept the backprimed fluid.
Before disconnecting a syringe from the cassette, pull up the
plunger slightly to avoid spilling the fluid. For rigid containers
(e.g., vials), open the cassette door, remove and invert the
cassette (ports down). Close upper slide clamp before
removing the container (to minimize spilling of fluid during
replacement of the container).
Содержание Plum XL
Страница 1: ...1 1 Rev...
Страница 2: ...System Operating Manual...
Страница 6: ...iv PlumXL NOTES...
Страница 8: ...2 Plum XL NOTES...
Страница 10: ...4 Plum XL NOTES...
Страница 16: ...10 Plum XL NOTES...
Страница 38: ...32 Plum XL NOTES...
Страница 42: ...36 Plum XL NOTES...
Страница 44: ...38 Plum XL NOTES...
Страница 46: ...40 Plum XL NOTES...
Страница 48: ...42 PlumXL NOTES...