System Operating Manual
25
7.0 TROUBLESHOOTING
This section contains solutions to routine clinical conditions
that may occur while using the Plum XL that do not require
assistance from hospital or Hospira Technical Service Center
personnel.
Problems that may occur in the Plum XL are in two categories:
alarms and malfunctions.
7.1
Alarms
During an alarm condition, the screen backlight and the alarm
message flash, and the audible alarm sounds. To clear an
alarm condition, proceed as follows:
Press the [SILENCE] key. Observe
the alarm message displayed. Turn
the control dial to HOLD/RESET.
Correct the alarm condition.
Turn control dial to RUN.
The following tips help correct the alarm conditions that may
occur:
MESSAGE
POSSIBLE CAUSE
CORRECTIVE ACTION
AIR IN LINE
Air detected distal
to cassette
Remove and reprime
cassette
AIR IN LINE
BACKPRIMING
Air detected
proximal to cassette
Container empty
Backprime to expel all
air
Change container and
backprime to expel air
OFF
CHARGE
SET RATE
SET VTBI
RUN
HOLD/RESET
CLEAR
VOL
OFF
CHARGE
SET RATE
SET VTBI
RUN
HOLD/RESET
CLEAR
VOL
Содержание Plum XL
Страница 1: ...1 1 Rev...
Страница 2: ...System Operating Manual...
Страница 6: ...iv PlumXL NOTES...
Страница 8: ...2 Plum XL NOTES...
Страница 10: ...4 Plum XL NOTES...
Страница 16: ...10 Plum XL NOTES...
Страница 38: ...32 Plum XL NOTES...
Страница 42: ...36 Plum XL NOTES...
Страница 44: ...38 Plum XL NOTES...
Страница 46: ...40 Plum XL NOTES...
Страница 48: ...42 PlumXL NOTES...