For Systems With Hospira MedNet
®
Software
140
System Operating Manual
Wireless Lan Upgrade Module
Battery Life:
With a new fully charged battery, the infuser
operates for a minimum of three hours at 125 mL/
hr or less, or delivers 250 mL if > 126 mL/hr. (Time
is measured from initial pumping to Depleted
Battery Alarm)
Recharge:
The battery charges whenever the infuser is
connected to AC power. The recharge time is
approximately six hours with the device operating
at 125 mL/hr on one line.
Electrical
Leakage:
Meets IEC 60601-1 standard: Medical Electronic
Equipment, Part 1: General Requirements for
Safety.
NURSE-CALL
System:
Circuitry
Ratings:
NURSE-CALL alarm is factory set for Normally-
Open (NO)
Contact the Technical Services Center to make an
internal adjustment to change the device from
Normally-Open (NO) to Normally Closed (NC)
system.
Voltage-30 VDC Max
Current- 0.25 Amps Max
Contact Rating- 3 Watts Max
Device Name:
Hospira MedNet 802.11 a/b/g Wireless
(Upgrade) Module
Standards:
IEEE802.11a/b/g
Transmit Power:
802.11 b/g- 17 dBm
802.11 a- 16 dBm
Antenna:
Integrated surface mount antenna
Certifications:
FCC Part 15.247, 15.407
IC RSS-210, RSS-102
This Device
Contains:
FCC ID: STJ80411396001
IC: 5627A- 80411396
Model: CUSTOM DWL-AG132