Plum A+
®
Infusion Systems
System Operating Manual
21
• The guidance included in this manual provides information
needed to:
• Determine the device’s suitability for use in the
intended environment.
• Manage the electromagnetic environment to permit the
device to perform as intended without disturbing other
equipment.
• Separate the device from all other electronic equipment. If
the device must be used near other electrical equipment,
monitor the equipment to ensure there is no electromagnetic
interference.
• Devices should not be used adjacent to or stacked with other
equipment. If the device must be used adjacent to or stacked
with other equipment, monitor the devices to verify normal
operation.
• USE ONLY components specifically labeled for use with the
Plum A+ Infusion System to help ensure the device operates
as intended.
• If you suspect external RF sources or other equipment are
influencing device operation, contact the biomedical
engineering department for additional guidelines concerning
electromagnetic immunity.
• Contact the biomedical engineering department for additional
information in the technical service manual concerning
operating devices near RF sources.
• The wireless module (20677/20791-04) has been tested with
Plum A+ infusion systems and has been found to comply with
the international standard IEC 60601-1-2 Edition 2
Electromagnetic Compatibility (EMC) of Medical Electrical
Equipment.