Plum A+
®
Infusion Systems
System Operating Manual
15
practicing epidural administration must be equipped with
resuscitative equipment, oxygen, naloxone, and other
resuscitative drugs. Adequate monitoring equipment (e.g.,
Oximetry) is recommended for continuous monitoring of the
patient during epidural administration. Patients must be
observed frequently for side effects in a fully-equipped and
staffed environment for at least 24 hours following completion
of drug administration by the epidural route. DELAYED
RESPIRATORY DEPRESSION FOLLOWING
CONTINUOUS EPIDURAL ADMINISTRATION OF
PRESERVATIVE-FREE MORPHINE SULFATE HAS BEEN
REPORTED.
• The epidural space has 58 openings through which fluid can
exit. Pressure buildup during administration is transient.
However, if a large volume of fluid is administered over a
short time period, the pressure will take longer to return to
normal. If overdelivery occurs during administration, observe
the patient closely for signs of spinal cord compression
(disorientation, headache, transient neuralgias) and drug
overdose.
B
ATTERY
O
PERATION
• When the battery is removed from the Plum A+, do not
operate on patients. Use of a properly maintained and
charged battery helps confirm proper operation.
• The battery may not be fully charged upon receipt. Connect
the infuser to AC power for at least six hours.
• Use AC power whenever possible. Connect to AC power
during storage to ensure a fully charged battery for
emergencies. If the quality of the earth grounding source is in
doubt, use battery power.
• If the low-battery alarm sounds, connect the infuser to AC
power immediately.
S
ETS
AND
A
CCESSORIES
• Only compatible LifeCare PlumSets
®
can be used with the
Plum A+. See individual set instructions for additional
information.
430-95597-004.book Page 15 Friday, March 9, 2007 8:30 AM