For Systems With Hospira MedNet
®
Software
20
System Operating Manual
• Maintaining a minimum separation distance of 2 ½ ft
between the Infusion pump system and portable/
mobile RF communications equipment
• List Numbers 12391 and 11971 are compliant to IEC/EN
60601-1-2 (1993)
List Numbers 20678, 20679, & 20792 are compliant to IEC/
EN 60601-1-2 (2001) and have been tested and found to
comply with EMC limits for the Medical Device Directive 93/
42/EEC (EN 55011 Class B and EN 60601-1-2:2001).
I
NTERCONNECTING
OF
M
EDICAL
E
QUIPMENT
• Accessory equipment connected to the analog and digital
interfaces must be certified according to the respective IEC
Standards (e.g. IEC 60950 for data processing equipment
and IEC 60601-1 for Medical Equipment). Furthermore, all
configurations shall comply with the system standard IEC
60601-1-1. Any person who connects additional equipment to
the signal input or output part configures a medical system,
and is therefore responsible for ensuring that the system
complies with the requirements of the system Standard IEC
60601-1-1. If in doubt, consult the technical service
department or your local representative.
Guidance on EMC Compatibility
• There is a shared responsibility between manufacturers,
customers and users to ensure that Medical Equipment and
Systems are designed and operated as intended. Medical
electrical equipment needs special precautions regarding
electromagnetic compatibility and needs to be installed and
used according to the electromagnetic compatibility
information provided in this manual.
• The device is suitable for use in all establishments, including
domestic establishments. If extended operation during
power mains interruption is needed, use battery power.
• Always manage the electromagnetic environment.