![Hoshizaki KM-1601MRH/3 Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/hoshizaki/km-1601mrh-3/km-1601mrh-3_service-manual_2165394029.webp)
29
IV. Service Diagnosis
WARNING
1. This unit should be diagnosed and repaired only by qualified service
personnel to reduce the risk of death, electric shock, serious injury, or fire.
2. Risk of electric shock. Use extreme caution and exercise safe electrical
practices.
3. Moving parts (e.g., fan blade) can crush and cut. Keep hands clear.
4.
CHOKING HAZARD: Ensure all components, fasteners, and thumbscrews
are securely in place after the unit is serviced. Make sure that none have
fallen into the storage bin.
5. Make sure all food zones in the icemaker and storage bin are clean after the
unit is serviced. For cleaning procedures, see "VI. Cleaning and Maintenance
Instructions."
A. Diagnostic Procedure
The diagnostic procedure is basically a sequence check which can be used at unit start-
up or for system diagnosis. Using this check out procedure will allow you to diagnose
electrical system and component failures in approximately 10 minutes under normal
operating conditions of 70°F or warmer air and 50°F or warmer water temperatures.
Before conducting a 10 minute checkout, check for correct installation, proper voltage per
unit nameplate and adequate water supply. As you go through the procedure, check to
assure the components energize and de-energize correctly. If not, those components and
controls are suspect. Check for voltage at the 10-pin connector.
1) Turn off the power supply, then access the control box. Clear any ice from bin control
bulb.
2) Turn on the power supply, then move the control switch to the "ICE" position. A 5 second
delay occurs.
3)
One Minute Fill Cycle – The inlet water valve is energized. After 1 minute, the control
board checks the float switch. If the float switch is closed, the unit cycles to harvest. If
closed, continue to step 4. If the float switch is open, the unit repeats the 1 minute fill
cycle until water enters and the float switch closes (low water safety protection during
initial start up and at the end of each harvest).
Diagnosis: If the water valve does not
open, check for no supply voltage at water valve terminals, bad coil, or plugged screen
or external filter (no water flow). If unit fails to start harvest, check for open float switch
or bad 1 minute timer in control board.