93-0061-00 Rev. A
7
Adflo
®
Turbo Unit
NOTE: If the Adflo
®
Turbo Unit starts to run during
charging, turn it off.
The Adflo
®
unit has a three-bar battery indicator (see
Figure 4) that is operational when the unit is turned on.
When all three bars are illuminated, the battery has 80-
100% capacity. Two bars indicate 20-80% capacity.
One bar indicates 5-20% capacity. During the last 30 to
60 minutes of charge, the bottom bar will flash
continuously one second ON then one second OFF and
the audible warning (beep) will beep four times in two
seconds, repeating every 30 seconds. The final 15 to 60
seconds causes the beeper to beep every 1/2 second.
NOTE: all airflow, lights and beeping will completely
stop when the battery is fully drained.
2.2 Pre-Operational Adflo
®
Turbo Unit
Inspection & Testing
Prior to beginning each work shift, perform the following
Pre-Operational Inspection to ensure the proper airflow
and alarm function of the Adflo
®
Turbo Unit.
1) Replace any missing or damaged parts.
2) Ensure that the HE filter is properly installed in the
Adflo
®
Turbo Unit. If using the OV/SD/CL/HC
cartridge, ensure that it is also properly installed in
the Adflo
®
unit. See Chapter 3 for instructions on
removing and replacing the HE filter and the
OV/SD/CL/HC cartridge.
3) Confirm that the battery pack is fully charged and
properly fitted to the Adflo
®
unit. See Chapter 3 for
instructions on removing and reinstalling the battery
pack.
4) If not already disconnected, disconnect the breathing
tube from the headpiece assembly’s airduct by
carefully twisting the breathing tube end connector
so it unlocks from the “J” channel on the airduct.
5) Insert the airflow indicator onto the end of the
breathing tube.
6) With the airflow indicator in a vertical position, turn
the Adflo
®
Turbo Unit on (see Figure 5). All three
LEDs located on the ON/OFF switch will turn on
for approximately one second; then the two green
LEDs should remain on.
7) Ensure that the bottom of the floating ball rests at, or
above, the minimum flow mark (see Figure 6).
NOTE: If the floating ball of the airflow indicator fails
to rest at or above the minimum flow level, refer to
Section 2.5 Fault Location of this manual.
Figure 4:
The Four Different Charge Indicators.
Figure 5:
Turning On the Adflo Unit.
Figure 6:
Airflow Indicator Check.