- 1 - 

1. Outline 

MV Series vaporizers have flow rate adjusting valves and vaporizer functions.    When one of 
them is combined with an appropriate model of LF or XF Series micro liquid mass flow 
meters for feedback control of the liquid flow rate, the equipment ensures stable generation 
of material.    Since the vaporizer unit [vaporizer]and the flow rate control unit[mixing valve] 
of each model of MV series are separated, it is possible to make independent temperature 
control. It is possible to set temerature of the vaporizer unit[vaporizer] higher than control 
unit[mixing valve], restraining the temperature rise in the flow control unit. It preuents heat 
decomposition in the flow rate control unit and enables stable supply. 

2. Specifications 

Model MV-2

□□□

-

□□□

 

Applicable material

*1 

TEOS etc. 

Heater specifications

*2 

MV-21

□□

-

□□□

: Mixing valve AC 120 V, 50/60 Hz, 200 VA, Vaporizer AC 120 V, 50/60 Hz, 200 VA 

MV-22

□□

-

□□□

: Mixing valve AC 208 V, 50/60 Hz, 200 VA, Vaporizer AC 208 V, 50/60 Hz, 200 VA 

MV-23

□□

-

□□□

: Mixing valve AC 240 V, 50/60 Hz, 200 VA, Vaporizer AC 240 V, 50/60 Hz, 200 VA

 

Orifice

 

MV-2

4

-

□□□

  :

0.4 mm 

MV-2

7

-

□□□

  :

0.7 mm 

MV-2

0

-

□□□

  :

1.0 mm 

 
Applicable flow rate

*3 

MV-2

□□

1-

□□□

  :Less than 1 ccm(or g/min) 

MV-2

□□

2-

□□□

  :1 to less than 5 ccm(or g/min) 

MV-2

□□

3-

□□□

  :5 to less than 10 ccm(or g/min) 

MV-2

□□

4-

□□□

  :10 ccm(or g/min) or more 

Wetted material

*4 

MV-2

□□□

-M

□□

  :SUS316L 

MV-2

□□□

-P

□□

  :SUS316L, PFA: Standard 

Pneumatic valve 

MV-2

□□□

-

V

: With pneumatic valve

 

*7 

MV-2

□□□

-

N

: Without pneumatic valve 

Filter MV-2

□□□

-

□□

F: With Filter 

MV-2

□□□

-

□□

N: Without Filter 

Temperature setting

*5

 

Depend on vaporization condition 

Temperature sensor 

K type thermocouple (Mixing valve unit and vaporizer unit) 

Temperature switches 

*6 

Mixing valve: For prevention of temperature rise:    UCHIYA EP42P 140

C NC type 

Vaporizer: For prevention of temperature rise: UCHIYA EP42P 140

C NC type 

External leak standard 

1

10

-8

 Pa

m

3

/s (He) or less 

Valve seat leak standard 

Mixing valve : 1

10

-6

 Pa

m

3

/s (He) or less [at working temperature] 

Pneumatic valve: 1

10

-9

 Pa

m

3

/s (He) or less [at working temperature] 

 
Standard joint 

Liquid inlet 

: 1/8 inch VCR type Male 

Carrier gas inlet: 1/4 inch VCR type Female 
Gas outlet 

: 1/2 inch VCR type Male 

Ambient temp. in main 
unit installation location 

15

C - 50

C    (altitude: up to 2000m) 

Pressure resistance 

1.0 MPa (G) 

Weight 

1,290 +/- 10 g (with pneumatic valve) 

1,150 +/- 10g (without pneumatic valve) 

 

Содержание MV-2000

Страница 1: ...Mixing Vaporizer MV 2000 Instruction Manual CODE I031170900...

Страница 2: ...arranties set forth herein shall be extended Any malfunction or damage attributable to improper operation Any malfunction attributable to repair or modification by any person not authorized by HORIBA...

Страница 3: ...to abide by the information in a WARNING may result in serious injury and can be life threatening WARNING CAUTION HOT SURFACE After heating up do not touch the main unit Doing so can burn your hands C...

Страница 4: ...iquid supply line 12 9 2 Clogging by vaporization 13 9 3 Pressure loss of internal nozzles 14 9 4 Valve travel for set temperature 14 9 5 Purging procedures 14 9 6 Temperature control 14 9 7 Setting o...

Страница 5: ...Hz 200 VA Orifice MV 2 4 0 4 mm MV 2 7 0 7 mm MV 2 0 1 0 mm Applicable flow rate 3 MV 2 1 Less than 1 ccm or g min MV 2 2 1 to less than 5 ccm or g min MV 2 3 5 to less than 10 ccm or g min MV 2 4 10...

Страница 6: ...differ in the material of the seat of the flow rate control valve The P PFA type valves are standard because seat leakage owing to particles and product hardly occurs For some materials having high p...

Страница 7: ...3 3 External Dimensions 3 1 without pnuematic valve Unit mm...

Страница 8: ...4 3 2 with pnuematic valve Unit mm...

Страница 9: ...rmine the piping diameter in consideration of piping pressure loss When the liquid material contains particles fit a filter 0 01 m or so in the middle of the line Please select the filter in considera...

Страница 10: ...e CA thermocouples and 200W heaters are used Use temperature regulators and SSRs appropriate to these specifications For MV use temperature regulators having a control period proportional period of 1...

Страница 11: ...ial check that the signal has been input When the liquid material is introduced the output signal of the Liqud flow meter LF or XF temporarily increases This is not an error The signal will become zer...

Страница 12: ...turn off the temperature regulators of MV Disconnect them after the main unit cools off 7 Troubleshooting When any trouble occurs with MV flow control or temperature condition check the following poi...

Страница 13: ...shut off after a temperature fall and a heater power supply when temperature does not rise and you check heater resistance and there is no electrical connection of a heater The heater power supply sh...

Страница 14: ...ing signal to the liquid flow meter LF or XF and check for seat leakage See the remedies stated in the following item Seat leakage The valve cable has not been connected to MV or Liquid flow meter LF...

Страница 15: ...e vaporizer of MV is controlled by the temperature regulator Therefore thermocouple and heater wires are necessary for each regulator Wire the regulator correctly checking the marking on the cables If...

Страница 16: ...ed and turned into air bubbles in the liquid The amount of gas dissolved in the liquid material changes depending on the applied pressure kind of pressure gas kind of liquid material and temperature I...

Страница 17: ...hanges It is necessary to compare the pressure under uniform conditions 2 The valve travel can be monitored by checking the supply voltage to the control valve valve voltage monitor output Since MV is...

Страница 18: ...the piping the material may not be purged completely Therefore for a material having an excessively low vapor pressure it is recommended to provide a branch line for evacuation on the piping for supp...

Страница 19: ...to Item 1 of 9 2 Clogging by vaporization and stop the vaporization before the secondary side is completely blocked If the passage is clogged to stop the flow the values output from the mass flow cont...

Страница 20: ...ll 5 Amp specification of overcurrent protection such as circuit breaker or fuse in the heater power line 1 Temperature regulation in mixing valve MV main unit 2 Temperature regulation in Vaporizer Mi...

Страница 21: ...ate of manufacture as far as the safety and usage precautions for the product are observed Name and amount of toxic hazardous substance or element used in a product Unit name Toxic hazardous substance...

Отзывы: