Chapter 2 Before Measurement (Meter SET)
16
HORIBA
Each HOLD condition is described below.
2.3 Custom Setting
Stability condition
Function
Auto hold
In the AUTO HOLD mode, the instrument judges potential stability
automatically to set the measurement values.
Manual hold
Determine an end point manually.
(Tap
to hold it.
)
1.
Select the CUSTOMIZE of the Hold type to set
the stability condition time and the stability
condition value.
2.
Use the numeric-key screen to enter
measurement variations as HOLD criteria for
each measurement item.
Tap
to return to the previous screen.
Mode
Measuring
target
Content
Time (s)
Temperature
(
C)
Criteria
【
Default
】
EXACT
COND, Resist
10
2.0
Minimum display digit: 1 digit
SAL
0.30 ppt (0.03%)
TDS
10 mg/L
NORMAL
COND, Resist
10
2.0
Minimum display digit: 3 digits
Default setting of
auto hold
SAL
1.00 ppt (0.10%)
TDS
30 mg/L
BRIEF
COND, Resist
10
2.0
Minimum display digit: 5 digits
SAL
3.00 ppt (0.30%)
TDS
100 mg/L
TIME
Common
-
-
Arbitrarily set at 2 s to 999 s.
【
10 s
】
CUSTOMI
ZE
COND
Arbitrary
setting
2 s to 60 s
【
10 s
】
2.0
Arbitrarily set at 0.001 mS/cm to
0.100 mS/cm (0.1 mS/m to 10.00
mS/m).
【
0.001 mS/cm
(0.1 mS/m)
】
SAL
Arbitrarily set at 0.10 PPT to
10.00 PPT (0.01% to 1.00%).
【
0.3 PPT
(0.03%)
】
Resist
Setting value of COND is reflected.
TDS
Arbitrarily set at 0.1 mg/L to 100
mg/L.
【
0.1 mg/L
】
We will explain the procedures of CUSTOMIZE
setting taking the AUTO HOLD item as an
example.