Measurement
29
Relay 1
Relay 2
Set ontact output
types ("r1","r2")
"non"
"AL"
"HoLd"
"CLn"
"nor"
Set contact output type (Type).
z
"non"; No output
z
"AL": Enables an alarm triggering
contact.
z
"HoLd": Outputs a contact during
transmission output Hold.
z
"CLn": Outputs a cleaning trigger or
a cleaning end trigger.
Set a control
method
(when the
contact output
type is set to
"AL".)
"Hi"
"Lo"
"Hi"
Specify whether control is performed by
upper limit operation or lower limit
operation.
z
"Hi": control by the upper limit
operation
z
"Lo": controlled by lower limit
operation.
Set an alarm
triggering value
(when the
contact output
type is set to
"AL".)
0 to
20000
4000
Define a alarm triggering value.
Press the ENT key to make changes
possible, make adjustments with the
▲
/
▼
key, and then press the ENT key to
confirm the changes made.
Set contact
output delay time
(when a contact
output type is set
to "AL" or
"Hold").
0 to 600
0
Define the delay time (in seconds) from
the time an alarm is raised until a
contact is output.
If the alarm contact point is released,
an alarm will be immediately released
without delay.
Set a cleaning
output type
(when a contact
output type is set
to "CLn").
"CLn"
"trg"
"CLn"
Specify a cleaning output type.
z
"CLn": Outputs a cleaning output
type in the HOLD state holding
during and after cleaning.
z
"trg": Outputs a contract for 5 s
within 2 s after the end of cleaning.
Set the FAIL
output at the
time of
abnormality
detection by self-
check.
"yES"
"no"
"yES"
Specify whether to enable FAIL output
when an error is judged as an
abnormality by the self-check function.
(E-80 to E-84)
Set the FAIL
output at the full-
scale over of a
measured value
"yES"
"no"
"no"
Specify whether to enable output FAIL
when a measured value exceeds
20000 mg/L.
(or 22000 mg/L if range cut is disabled )
Set the FAIL
output delay time
0 to 600
0
Define the delay time (in seconds) from
the time FAIL is triggered until the
contact point output is performed.
If a FAIL contact point is released, FAIL
is immediately released.
Screen 1
Screen 2
Description
Setting
range
Initial
value
Remarks