Overview
7
Description of operation modes and menu items
This equipment offers three types of operation modes and four types of menus.
Q
Operation mode
Meas mode
In this mode, samples are measured, and transmission signals and relays are controlled.
Cal mode
This mode allows you to perform zero calibration and span calibration.
The Cal menu is also used to perform calibration.
Hold mode
This mode allows you to stop measurements to do the processing in the Setup menu, the Cal
menu or the User Check menu.
Q
Menu
The Meas mode and the Hold mode have the following menus in the lower tier.
For selecting each mode or item, see the corresponding paragraph.
Meas mode
Cal/Control Value Setting menu
This menu allows you to check the calibrated value or to set/change some controlled values
while checking the output.
Hold mode
Setting menu (SEt)
Prior to operation, this menu allows you to set all the measurement-related parameters, such
as the detector information input and the allocation output.
Cal menu (CAL)
This menu allows you to perform zero calibration and span calibration of the MLSS detector.
User Check menu (USr)
This menu allows you to check the output status, the measured value, and other data. The set
values can be reset to the factory settings.