Reference data
10
Reference data
Specifications
Product name
pH meter
Model
HP-100
Combined sensor
Glass electrode
Measurement
range
pH0 to pH14
Display resolution
0.01pH
Performance
Repeatability
Within
±
0.05 pH (response for equivalent input)
Linearity
Within
±
0.05 pH (response for equivalent input)
Analog output
Output type
4 mA to 20 mA DC: input/output isolated type
Load resistance
Maximum: 500
Ω
Case of 24 V DC power supply
Linearity
Within
±
0.08 mA (output only)
Repeatability
Within
±
0.08 mA (output only)
Output range
pH0 to pH14
Temperature
compensation
Applicable temperature
element
Platinum resistor: 1 k
Ω
(0
°
C)/6.8 k
Ω
(25
°
C) Optionally available
Temperature
compensation range
0
°
C to 100
°
C
Calibration
Calibration method
Manual calibration
Number of calibration
points
Selectable 2 points
Operating temperature
range
−
10
°
C to 50
°
C (without freeze)
Operating humidity
range
Relative humidity: 5
%
to 90
%
(without condensation)
Storage temperature
−
25
°
C to 65
°
C
Power supply
Power supply voltage
range
24 V DC (22.5 V to 30 V)
Power consumption
0.6 W (max)
Compatible standards
CE
marking
Compatible
standards
EMC Directive (2004/108/EC) EN61326-1:2006
EMC
Immunity
Industrial location
Electrostatic discharge
IEC61000-4-2
Electromagnetic field of radiated radio
frequency
IEC61000-4-3(*Note 1)
Electric fast transient/burst
IEC61000-4-4
Surge
IEC61000-4-5(*Note 2)
Conducted interference induced by radio
frequency
IEC61000-4-6(*Note 1)
Emission
ClassA
Radiated disturbance
CISPR 11 CLASS A
FCC Rules
Part15 CLASS A