Set Up
3
External Wiring
Make wiring in accordance with the external wiring diagram. Use the optional janction box (JB-
30) to relay between the electrode and transmitter.
Use optional special (W-5B). Maximum length of wiring should be within 50 m.
Note
z
Class D grounding.
z
Use a transmission cable equivalent to CVVS 2 mm
2
.
z
To connect the terminal, use M4 caulking terminal.
z
Supply the power after checking connection.
WARNING
ELECTRIC SHOCK
Opening the rear cover may cause electric shock while supplying electric power. Do not open the rear
cover without disconenecting the main power.
ELECTRIC SHOCK
Maintain ground to avoid electric shock. Do not ground the unit to a gas pipe or to an indirect ground,
such as another piece of the equipment.
CAUTION
FIRE
Do not put water, metal chip or any foreign substances inside the rear cover. The unit may be shorted
catch fire.