24
The power is not
turned ON.
Check that the batteries are inserted properly. If the batteries
are low, replace them both with new ones at the same time.
Er1 is displayed
right after power
ON.
The internal IC in the meter may defect.
After Er1 is displayed, the meter enters the initialization mode
automatically and Init and
appear. Press the CAL switch for
over 2 seconds to execute initialization, and then turn OFF and
ON again (refer to page 22).
If Er1 is still displayed after the initialization, the internal IC in
the meter defects. Replace the product with new one (the
meter can not be repaired).
Er2 is displayed
right after power
ON.
The internal IC in the meter defects. Replace the product with
new one (the meter can not be repaired).
Er3 is displayed
right after power
ON.
The internal IC in the meter defects. Replace the product with
new one (the meter can not be repaired).
How can I return all
the settings of the
special setting mode
to the default
settings?
Perform initialization (refer to page 22).
Question
Answer