horiba B-711 Скачать руководство пользователя страница 31

23

8

Appendix

8.1

Frequently asked questions

Question

Answer

How long is the 
sensor's service 
life?

It depends on measurement samples and conditions.
It should be approx. 1500 measurements for typical samples.
Note that sensor deterioration and failure are not included in 
the warranty.

How can I check the 
sensor's condition?

For B-712 and B-713, perform two-point calibration. If 
calibration error occurs, the sensor is deteriorated. Replace the 
sensor.
For B-711, measure the pH 4.01 standard solution (part No. 
3200457726) sold seperately after calibration, and check how 
much the measured pH value drifts from the specified pH. if the 
measured value drifts, the sensor is deteriorated. Replace the 
sensor.

Are there any helpful 
tips or precautions 
to be aware for 
measurement?

Use the light shield cover to avoid direct sunlight during 
measurement because the sensor is affected by light.

When the sample amount is enough, washing the sensor twice 
or so with the sample allows more accurate measurement.

Residue between the light shield cover and flat sensor pre-
vents accurate measurement. Before measurement of the next 
sample, clean the sensor with tap water and remove moisture. 

Can I measure high- 
or low-temperature 
samples?

This product can not measure a sample with a temperature 
outside the meter's operating temperature range (5

q

C to 40

q

C).

The difference between the sample temperature and ambient 
temperature increases the measurement error. Perform mea-
surement after the sample reaches the ambient temperature. 

The measured value 
does not change 
after changing the 
sample.

If 

 lights steadily, the measured value is locked.

Press the MEAS switch to unlock the value. 
If the value does not change after unlocking, the sensor may 
be damaged. Replace the sensor.

“0.0” blinks in pH 
measurement.

The measured pH may be out of the specified measurement 
range (2 pH to12 pH). Measure a standard solution to check, 
and if “0.0” blinks even then, replace the sensor.

The temperature 
alarm icon blinks 
during 
measurement.

The measuring environment temperature may not meet the 
specified operating temperature (5

q

C to 40

q

C).

When the environment temperature is within the specified 
range and the alarm icon blinks, replace the sensor.

Содержание B-711

Страница 1: ...CODE GZ0000297059...

Страница 2: ......

Страница 3: ...Refer to page 6 to page 22 for detailed instructions...

Страница 4: ...h herein shall be extended z Any malfunction or damage attributable to improper operation z Any malfunction attributable to repair or modification by any person not authorized by HORIBA Ltd z Any malf...

Страница 5: ...tive 2006 66 EC in the European Union The symbol might be put with one of the chemical symbols below In this case it satisfies the requirements of the Directive 2006 66 EC for the object chemical This...

Страница 6: ...ble protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and use...

Страница 7: ...ration at pH 7 B 711 B 712 B 713 10 5 3 One point calibration at pH 4 B 712 B 713 10 5 4 Two point calibration B 712 B 713 11 6 Measurement 12 6 1 Sample setting 12 6 2 Measurement operations 15 7 Spe...

Страница 8: ......

Страница 9: ...tions pH 4 pH 4 01 1 1 pH 7 pH 6 86 1 1 pH 7 pH 7 00 1 Pipette 1 1 1 Sampling sheet B 5 sheets 1 1 1 Instruction manual Common to all model 1 1 1 Quick start manual 1 1 1 Part No Name Type Application...

Страница 10: ...the sampling sheet holder cover refer to page 14 shield the flat sensor from light with an alternative 3 Lithium batteries CR2032 u2 4 MEAS switch Switches the calibration mode to the measurement mode...

Страница 11: ...ights up when the batteries are low and need to be changed 12 Temperature alarm icon Blinks when the measuring environment temperature does not meet the specified operating temperature 5qC to 40qC 13...

Страница 12: ...ed The waterproofing gasket is seated properly in the groove with no twisting or warping The meter and sensor are not deformed z Neither the meter nor sensor is waterproof by itself The sensor must be...

Страница 13: ...junction of the flat sensor The appearance of this powder or solution is normal Simply rinse it off with water before use J Battery z Keep batteries out of reach of children If someone accidentally sw...

Страница 14: ...ies at once Do not use old and new batteries together J Inserting the batteries 1 Slide both batteries into the battery case as shown Be sure to use two CR2032 batteries and put them with the plus sid...

Страница 15: ...z When removing the sensor do not let any water get inside the meter If some moisture remains on the waterproof gasket wipe it off carefully J Attaching the sensor 1 Confirm that the waterproofing gas...

Страница 16: ...h the flat sensor strongly It may damage the sensor 2 Close the light shield cover and the slide cap then store the meter J Power ON 1 Press and hold the ON OFF switch for over 2 seconds The power is...

Страница 17: ...ecommended especially if sample pH is outside 1 pH from the pH of the standard solution J Precautions for calibration z If remains blinking and Err error display appears the calibration is failed Chec...

Страница 18: ...the sensor with tap water and remove moisture 5 Press the MEAS switch for 0 5 seconds to enter the measurement mode and prepare for measurement 1 Set One point calibration at pH 4 refer to page 19 2 O...

Страница 19: ...lue is displayed After the calibration is completed and stop blinking and light steadily 4 After the calibration for pH 7 is completed open the light shield cover to remove the pH 7 standard solution...

Страница 20: ...e the moisture on it Tip J Drop 1 Open the light shield cover and put some drops of sample on the flat sensor to cover the entire flat sensor 2 Close the light shield cover Try to use the provided sam...

Страница 21: ...3 times J Scooping 1 Open the sliding cap on the light shield cover 2 Immerse the sensor into the sample and stir gently 2 or 3 times and then scoop up some of the sample with the sensor 3 Place the m...

Страница 22: ...ote The sampling sheet holder cover does not shield the sensor from light which affects the sensor When using the sampling sheet holder cover shield the flat sensor from light with an alternative 1 Re...

Страница 23: ...the sensor 2 Read the displayed value when appears J With using the reading locking function 1 Confirm that the meter is in the measurement mode and set a sample on the sensor 2 After appears press th...

Страница 24: ...bient temperature measured using an internal temperature sensor 7 2 Calibration point setting only for B 712 B 713 page 19 Used to select the calibration points 7 3 Display digit setting only for B 71...

Страница 25: ...17 7 6 Software version display mode page 22 The current software version is displayed Returns to the display for temperature display mode CAL switch...

Страница 26: ...s a guide 1 Press and hold the MEAS switch for over 3 seconds in the measurement mode to enter the special setting mode All items appear on the LCD and then the dis play changes as shown right 2 Press...

Страница 27: ...urned 1 Press and hold the MEAS switch for over 3 seconds in the measurement mode to enter the special setting mode All items appear on the LCD and then the dis play changes as shown right 2 Press the...

Страница 28: ...the MEAS switch to apply the setting The measurement mode is returned 1 Press and hold the MEAS switch for over 3 seconds in the measurement mode to enter the special setting mode All items appear on...

Страница 29: ...curve 3 Press the MEAS switch for 0 5 seconds The voltage output from the sensor is displayed 4 Press the MEAS switch to return the measurement mode 1 Press and hold the MEAS switch for over 3 second...

Страница 30: ...wn right 2 Press the CAL switch until Init appears 3 Press the MEAS switch for 0 5 seconds appears 4 Press the CAL switch for over 2 seconds All settings and all calibration data are reset to the fact...

Страница 31: ...between the light shield cover and flat sensor pre vents accurate measurement Before measurement of the next sample clean the sensor with tap water and remove moisture Can I measure high or low tempe...

Страница 32: ...n OFF and ON again refer to page 22 If Er1 is still displayed after the initialization the internal IC in the meter defects Replace the product with new one the meter can not be repaired Er2 is displa...

Страница 33: ...ours in continuous use Main materials ABS epoxy Outer dimensions mass 164 mm 29 mm 20 mm excluding projections Approx 50 g meter only without batteries Main functions Temperature compensation waterpro...

Страница 34: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...

Отзывы: