Ion concentration measurement
48
When calibrating three points or more, use the same procedure to continue calibrating
the 3rd and subsequent points after the 2nd point calibration. You can calibrate up to five
points.
Note
• If calibration of any standard solution is performed again, only the value of calibrated
solution is updated. If calibration is repeated, only the calibration value of used standard
solution is updated. The calibration values are retained until next calibration is performed
or the data is deleted. For the procedure for deleting the calibration data, refer to
"Deleting calibration data" (page 61).
• It is recommended that you should clear the previous calibration data before the
calibration in order to perform more accurate calibration.
•Check of the ion electrode status
After calibration is complete, the current ion electrode status is diagnosed from the
calibration result.
Display
Description
Reference
ERR No. is not
displayed
Good condition.
-
ERR No.0005
Ion valency setting is incorrect or sensitivity setting is
out of the measurable range.
Set the correct ion valency. If the setting is correct but
an error still occurs, perform electrode maintenance or
replace the electrode.