Ion concentration measurement
42
To return to the measurement mode, press the
key.
Note
• Be sure to set the unit before performing calibration, as the calibration calculation varies
depending on the unit and ion valency. If not, incorrect measurement values may result.
After changing the unit or ion valency of the sample, change the setting as appropriate for
the ion valency of the sample and recalibrate prior to performing measurement.
• When the unit of ion concentration is changed, the previous calibration data is deleted.
key
key
key
key
key
key
(Ex.: "mg/L" is selected)