Ion concentration measurement
46
3. Perform the 1st point calibration. Immerse at
least 3 cm from the tip of the ion electrode in
the standard solution.
Make sure that the responsive membrane is
immersed.
4. Press the
key to enter the calibration
mode.
Stability judgment starts and the ION icon blinks
until the measurement value is stabilized. When
the value is stabilized, the ION icon changes
from the blinking state to the lit state.
5. Press the
keys to enter the ion concentration value of the standard
solution used for calibration.
To change the decimal point, press the
key.
Changing unit of calibration value is available at the ion unit parameter in the SET
mode.
Note
The ion concentration value of the standard solution can be entered even when "Or" is
displayed on the screen.
key
Stabilized
key
Enter a
number