Troubleshooting
94
●
Nothing appears when the power is turned ON
●
Swelling of operation key sheet
●
Part of the display is missing
Cause
How to solve problem
Power is not supplied.
Insert batteries or connect the AC adapter (option).
Battery polarity (+,
−
) is
reversed.
Insert the batteries with the polarity (+,
−
) correctly
oriented.
Battery life is low.
Replace the batteries or connect the AC adapter (option).
Instrument defect
Consult your dealer.
Cause
How to solve problem
Using the instrument at high
elevation or other location where
the air pressure is different from
sea level.
To eliminate the pressure difference between the inside
and outside of the instrument, briefly open and then close
the AC adapter cover. After opening, correctly close the
cover to maintain dust and water proofing.
Instrument defect
Consult your dealer.
Cause
How to solve problem
Instrument defect
Check the display in full screen state when turning ON
the power of the instrument.