When using the 3-Point Known Load Calibration type, the following menu items
are available:
•
“
KNOWN POINT 1/3
”: point 1 of 3, usually 0% of the transducer’s full-scale capac-
ity.
•
“
KNOWN POINT 2/3
”: point 2 of 3, usually 50% of the transducer’s full-scale capac-
ity.
•
“
KNOWN POINT 3/3
”: point 3 of 3, usually 100% of the transducer’s full-scale
capacity.
When using the 5-Point Known Load Calibration type, the following menu items
are available:
•
“
KNOWN POINT 1/5
”: point 1 of 5, usually 0% of the transducer’s full-scale capac-
ity.
•
“
KNOWN POINT 2/5
”: point 2 of 5, usually 25% of the transducer’s full-scale capac-
ity.
•
“
KNOWN POINT 3/5
”: point 3 of 5, usually 50% of the transducer’s full-scale capac-
ity.
•
“
KNOWN POINT 4/5
”: point 4 of 5, usually 75% of the transducer’s full-scale capac-
ity.
•
“
KNOWN POINT 5/5
”: point 5 of 5, usually 100% of the transducer’s full-scale
capacity.
These menu items are only available with the Known-Load Calibration types.
NOTICE
To insure both correct operation of the transducer and application of the load,
the instrument expects the voltage applied at each known-load point to be
increasing. For example, the load applied at Known-Load Point 2/2 must
cause the transducer to produce a more positive voltage than at Known-Load
Point 1/2.
Содержание SC500
Страница 8: ...page 8 008 0671 00 ...
Страница 14: ...page 14 008 0671 00 ...
Страница 18: ...page 18 008 0671 00 ...
Страница 24: ...page 24 008 0671 00 ...
Страница 38: ...page 38 008 0671 00 ...
Страница 42: ...page 42 008 0671 00 ...
Страница 50: ...page 50 008 0671 00 ...
Страница 72: ...page 72 008 0671 00 ...
Страница 82: ...page 82 008 0671 00 ...
Страница 90: ...page 90 008 0671 00 ...
Страница 116: ...page 116 008 0671 00 ...
Страница 126: ...page 126 008 0671 00 ...