page 128 008-0671-00
15.2 Error Message List
Error 05,
DIVIDE BY ZERO
A Mathematics Virtual channel has attempted to divide by zero.
Error 07,
SQUARE ROOT
A Mathematics Virtual channel has attempted to take the square root of a negative
number.
Error 09,
LOG/LN
A Mathematics Virtual channel has attempted to take the logarithm of either zero or a
negative number.
Error 16,
+15V SUPPLY HIGH
The in15 VDC power supply is generating a voltage that is too high.
Error 17,
+15V SUPPLY LOW
The in15 VDC power supply is generating a voltage that is too low. Check the
wiring between the instrument and the transducer or amplifier.
Error 18,
-15V SUPPLY HIGH
The internal -15 VDC power supply is generating a voltage that is too high (more pos-
itive). Check the wiring between the instrument and the transducer or amplifier.
Error 19,
-15V SUPPLY LOW
The internal -15 VDC power supply is generating a voltage that is too low (more nega-
tive).
Error 20,
CHANNEL >MAXCH
An invalid channel number has been encountered by the instrument. Restart the
instrument. If this error persists, contact our Service department with the information
found in the
SYSTEM MENU
->
SOFTWARE REV.
menu item.
Error 21,
BLOWN STACK
The instrument’s software stack has overflowed. Restart the instrument. If this error
persists, contact our Service department with the information found in the
SYSTEM MENU
-
>
SOFTWARE REV.
menu item.
Error 22,
TYPE NOT FOUND
The instrument has not been programmed to understand the type of channel. Contact
our Service department with the information found in the
SYSTEM MENU
->
SOFTWARE REV.
menu item.
Error 23,
CAN'T RENDER FS
The full-scale value of the transducer cannot be rendered on the display with the
selected
DISPLAY SETUP
menu settings. Change the
DISPLAY SETUP
menu settings to better
match the transducer’s capacity.
Error 24,
SIG.MOD CRC ROM
A checksum error occurred while reading a
Signature Module’s
read-only-memory.
Check the wiring and restart the instrument. This error may occur with very long cable
lengths.
Error 25,
SIG.MOD CRC READ
A checksum failure occurred reading data from a Signature Module. Check the wiring.
This error may be caused by a faulty
Signature Module
.
Error 26,
NO SIG.MOD.
No
Signature Module
was detected during
UPDATE SIG.MOD.
or
DEFAULT SIG.MOD.
Check
wiring.
Error 27,
S.EE WRITE NAK
The channel’s EEPROM could not be detected during a write operation. Restart the
instrument.
Error 28,
S.EE READ NAK
The channel’s EEPROM could not be detected during a read operation. Restart the
instrument.
Error 29,
S.EE VERIFY FAIL
A verification failure occurred during a write operation to the channel’s EEPROM.
Содержание SC500
Страница 8: ...page 8 008 0671 00 ...
Страница 14: ...page 14 008 0671 00 ...
Страница 18: ...page 18 008 0671 00 ...
Страница 24: ...page 24 008 0671 00 ...