l Inpu
t C
ha
nnel
12
12.6.3
AUXn FUNCTION
Menu
Items
The
AUX1 FUNCTION
and
AUX2 FUNCTION
menu items determine what happens when the
Auxiliary Function pins (labeled as “AUX1” and “AUX2”) on the Input Channel con-
nector are activated. These pins are “activated” when connected to the DGND pin.
The choices are:
•
“
DISABLED
” means that activating the pin does nothing.
•
“
TRACK HOLD
” means that the tracking, peak and valley values will not be
updated.
•
“
HIGH/LOW HOLD
” means that the peak and valley values will not be updated.
•
“
HIGH/LOW CLEAR
” means that the peak and valley values are reset.
•
“
TARE ON
” activates the channel’s Tare function (resets it to zero) .
•
“
TARE OFF
” deactivates the channel’s Tare function.
As the Auxiliary Function pins are not isolated, it is recommended that a push-but-
ton switch or relay is used to connect these pins to the DGND pin.
Figure 12-9: Auxilliary Input Example
NOTICE
A SensoCode program running on a Mathematics Virtual Channel may over-
ride the default behavior of the Auxilliary Function inputs. Consult the Customer
Information Sheet included with your instrument for details.
Содержание SC500
Страница 8: ...page 8 008 0671 00 ...
Страница 14: ...page 14 008 0671 00 ...
Страница 18: ...page 18 008 0671 00 ...
Страница 24: ...page 24 008 0671 00 ...
Страница 38: ...page 38 008 0671 00 ...
Страница 42: ...page 42 008 0671 00 ...
Страница 50: ...page 50 008 0671 00 ...
Страница 72: ...page 72 008 0671 00 ...
Страница 82: ...page 82 008 0671 00 ...
Страница 90: ...page 90 008 0671 00 ...
Страница 116: ...page 116 008 0671 00 ...
Страница 126: ...page 126 008 0671 00 ...