background image

ELSTER RABO

®

 Rotary Gas Meter

07   Honeywell

Pete’s Plugs

®

 are

 provided for simplified pressure connections 

[Figure  9].  Typically,  when  proving  a  meter,  both  the  meter 
pressure  and  meter  differential  pressure  are  used  and 
recorded. The RABO meter differential pressure ports are not 
full  bore  and  most  temperature  probes  will  not  fit  through 
into the gas stream. Installing the temperature probe in the 
provided thermowell ports is recommended.

b) Differential Pressure Testing

A differential pressure test is not an accuracy test, but it can 
be used to gauge the relative performance of the meter.
Rotary meters are made from solid parts, machined to tight 
tolerances with close clearance fits, and the energy it takes 
to turn the meter is generally used to overcome the friction 
of the bearings and other rotating parts. If this friction value 
increases,  it  will  take  more  energy  to  overcome  it.  This 
additional  energy  is  measured  as  an  increased  pressure 
drop across the meter.
The  increased  friction  is  caused  by  contamination  of  the 
measuring chamber and/or bearings.
Contamination  in  the  measuring  chamber  can  cause  the 
meter components to wear against each other. Contamination 
in the bearings makes them harder to turn.
Comparing the pressure drop reading when the meter is new 
to the reading collected after time allows the performance to 
be monitored.
RABO meters are supplied with 

Pete’s Plugs

®

 installed in the 

meter  differential  pressure  ports  [Figure  9].  This  facilitates 
differential  meter  testing.  A  differential  pressure  test  is 
performed using a differential pressure manometer. The test 
is to measure the pressure drop across the meter at a given 
flow rate, under known operating conditions at a known date 
and time. 
The  meter  differential  curve  of  flow  rate  vs  pressure  is  not 
linear.  Testing  at  multiple  flow  rates  is  suggested.  It  is 
preferable to test at 3 rates between 25% and 100% of flow 
if  possible.  Differential  pressure  tests  at  flow  rates  under 
25%  are  hard  to  interpret  because  the  meter  differential 
pressure at the lower flow rates is quite small. The error in 
measurement is almost as large as the reading itself.
The  meter  differential  pressure  is  also  a  function  of  line 
pressure  and  increases  as  line  pressure  increases.  Testing 
the  meter  at  the  same  conditions  (line  pressure  and  flow 
rates) yields comparable data. 
A  change  in  the  differential  pressure  indicates  a  change 
in  performance.  Testing  has  shown  that  a  50%  increase  in 
meter differential pressure (at flow rates over 25%), indicates 
almost a 1.0% change in meter accuracy. 
Baseline  data  must  be  captured  during  initial  start-up  to 
which future data can be compared.
If the differential pressure test shows an increase in the meter 
pressure drop at a given flow rate of more than 50% from the 
original value (1.5 x original value), then it is recommended 
that the meter be removed and serviced.
The  frequency  of  deferential  testing  is  at  the  discretion  of 
the user.

Figure 9. 

Pete’s Plugs

®

The meter volume is determined though a device that senses 
the meter revolutions. An optical pickup can be used on the 
index proving wheel or an index pulser may be used. 
On-site proving, performed in-line, also requires connections 
that  allow  air  flow  through  the  meter  to  the  prover.  This  is 
accomplished by use of pipe Tees upstream and downstream 
of the meter. A bypass loop is recommended for uninterrupted 
gas supply to the customer.
While  performing  an  accuracy  test,  the  meter  is  tested  for 
a specified volume of gas at each flow rate test point. Poor 
repeatability at a given rate may be caused by a test volume 
that is too small. It the meter does not repeat within acceptable 
limits  (0.1%),  try  increasing  the  test  volume  and  retesting 
the  meter.  Most  companies  have  developed  test  plans 
that  include  the  volume  of  gas  for  each  size  meter  at  a 
given  flow  rate.  Consult  your  Elster  representative  for 
applicable volumes. 

PETE’S PLUGS

®

RABO

®

 Rotary Gas Meter

06 Elster Instromet

Figure 9. Pressure test ports

Pressure  test  ports  are  provided  for  simplified  pressure 
connections [Figure 9]. 

Below  45

O

F  the  neoprene  core  of  the  Pressure  test  ports 

does  not  recover  its  original  shape  as  rapidly  as  it  would 
at  temperatures  above  45

O

F.  Therefore,  upon  removal  of  

the  probe  used  for  pressure  or  temperature  readings,  the 
valves  may  not  close  fully  and  immediately  or  the  valves  
may  remain  slightly  open  until  the  operating  temperature  
is  above  45

O

F.  Duration  of  probe  insertion  and  pressure  

are also factors in the rate of valve closing. For this reason, 
plugs should not be used in applications where discharging 
gas  or  liquids  would  create  a  hazard.  The  gasketed  cap  
is  supplied  to  eliminate  the  small  amount  of  leakage  that  
may occur at lower temperatures after probe insertion and 
removal.  The  cap  should  be  left  on  the  plug  at  all  times  
and  tightened  to  deter  unauthorized  removal.  Special  care 
should  be  taken  to  assure  that  readings  are  taken  in  the 
shortest  space  of  time  and  that  a  probe  is  never  left  in  a  
plug  for  a  period  of  hours  or  days.  Severe  deformation  of 
the  valves  may  occur  if  the  probe  is  left  in  the  plug  for  a  
long period of time.

b)   Differential Pressure Testing 

Typically, when proving a meter, both the meter pressure and 
meter differential pressure are used and recorded. The RABO 
meter differential pressure ports are not full bore and most 
temperature probes will not fit through into the gas stream. 
Installing the temperature probe in the provided thermowell 
ports is recommended.

The meter volume is determined though a device that senses 
the meter revolutions. An optical pickup can be used on the 
index proving wheel or an index pulser may be used. 

On-site proving, performed in-line, also requires connections 
that  allow  air  flow  through  the  meter  to  the  prover.  This  is 
accomplished by use of pipe Tees upstream and downstream 
of the meter. A bypass loop is recommended for uninterrupted 
gas supply to the customer. 

A differential pressure test is not an accuracy test, but it can 
be used to gauge the relative performance of the meter. 

Rotary meters are made from solid parts, machined to tight 
tolerances with close clearance fits, and the energy it takes 
to  turn  the  meter  is  generally  used  to  overcome  the  friction 
of the bearings and other rotating parts. If this friction value 
increases,  it  will  take  more  energy  to  overcome  it.  This 
additional  energy  is  measured  as  an  increased  pressure 
drop across the meter. 

The  increased  friction  is  caused  by  contamination  of  the 
measuring chamber and/or bearings. 

Contamination  in  the  measuring  chamber  can  cause  the  
meter components to wear against each other. Contamination 
in the bearings makes them harder to turn. 

Comparing the pressure drop reading when the meter is new 
to the reading collected after time allows the performance to 
be monitored. 

RABO meters are supplied with Pressure test ports

 

installed  

in  the  meter  differential  pressure  ports  [Figure  9].  This  faci- 
litates differential meter testing. A differential pressure test is 
performed using a differential pressure manometer. The test 
is to measure the pressure drop across the meter at a given 
flow rate, under known operating conditions at a known date 
and time. 

The  meter  differential  curve  of  flow  rate  vs  pressure  is  not 
linear.  Testing  at  multiple  flow  rates  is  suggested.  It  is 
preferable to test at 3 rates between 25% and 100% of flow 
if  possible.  Differential  pressure  tests  at  flow  rates  under  
25%  are  hard  to  interpret  because  the  meter  differential 
pressure  at  the  lower  flow  rates  is  quite  small.  The  errorin 
measurement is almost as large as the reading itself. 

The  meter  differential  pressure  is  also  a  function  of  line 
pressure  and  increases  as  line  pressure  increases.  Testing 
the  meter  at  the  same  conditions  (line  pressure  and  flow 
rates) yields comparable data. 

A  change  in  the  differential  pressure  indicates  a  change 
in  performance.  Testing  has  shown  that  a  50%  increase  in 
meter differential pressure (at flow rates over 25%), indicates 
almost a 1.0% change in meter accuracy. 

Baseline  data  must  be  captured  during  initial  start-up  to  
which future data can be compared. 

If the differential pressure test shows an increase in the meter 
pressure drop at a given flow rate of more than 50% from the 
original value (1.5 x original value), then it is recommended 
that the meter be removed and serviced. 

The  frequency  of  deferential  testing  is  at  the  discretion  of  
the user.

Figure 5. Horizontal installation

Figure 6. Vertical installation

Figure 7. Sight glass

Figure 4. Side view

b) Mounting

1)

Always  follow  your  company’s  procedures,  and 
applicable local codes and ordinances.

2)

Ensure gas valves are closed.

3)

Ensure  the  upstream  piping  is  clean  and  free  of  any 
debris.

4)

Remove protective caps from meter inlet and outlet prior 
to installation.

5)

Ensure the impellers turn freely.

6)

Ensure  the  direction  of  flow  using  the  arrow  on  the 
nameplate.

7)

Ensure  the  meter  orientation  is  correct.  Impeller  shafts 
must be horizontal [Figure 4].

8)

Connect  the  inlet  and  outlet  pipe  flanges  using 
appropriate  bolts  and  gaskets.  Inlet  and  outlet  pipe 
flanges should be parallel and should not introduce any 
bind on the meter body when tightened.

9)

Level meter to within 1/16" per foot in all directions and 
tighten flange bolts evenly (maximum 80ft-lbs). 

c) Adding Oil

!

WARNING

Add oil only to the index end of the meter.

1)

Ensure gas valves are closed and meter and piping are 
depressurized.

!

WARNING

Failure  to  depressurize  the  meter  prior  to  removing 

meter  and/or  components  could  result  in  personal 

injury and/or property damage.

2)

Remove oil fill plug in the counter end case cover using a 
5mm hex key [Figures 5 and 6].

3)

Using the supplied syringe and oil, slowly add oil until 
it is +/-1/16" of the center of the sight glass [Figure 7]. 

DO NOT OVERFILL.

 Only use Shell Morlina lubricating oil.

!

WARNING

DO NOT remove any sight glasses. No maintenance 

can be performed through these openings.

4)

Reinstall the oil fill plug.

OIL

FILL

OIL

LEVEL

SIGHT

GLASS

OIL

DRAIN

OIL

FILL

OIL

FILL

OIL

FILL

OIL

LEVEL

SIGHT

GLASS

OIL

DRAIN

RABO

®

 Rotary Gas Meter

04 Elster Instromet

While  performing  an  accuracy  test,  the  meter  is  tested  for 
a specified volume of gas at each flow rate test point. Poor 
repeatability at a given rate may be caused by a test volume 
that is too small. It the meter does not repeat within acceptable 
limits  (0.1%),  try  increasing  the  test  volume  and  retesting  
the  meter.  Most  companies  have  developed  test  plans  
that  include  the  volume  of  gas  for  each  size  meter  at  a 
given  flow  rate.  Consult  your  Elster  representative  for 
applicable volumes.

PRESSURE TEST PORTS 

Содержание RABO

Страница 1: ...Instromet RABO Rotary Gas Meter Instruction Manual ...

Страница 2: ......

Страница 3: ...15 Troubleshooting 11 RABO Rotary Gas Meter RABO Rotary Gas Meter 02 Elster Instromet RABO Rotary Gas Meter 1 General Information 3 2 Intended Use and Application 3 3 Operating Principle 3 4 Receiving Handling and Storage 4 5 Installation 4 6 Start up Commissioning 6 7 Decommissioning and Removal 6 8 Inspection and Maintenance 6 9 Testing 6 10 Technical Data 8 11 Index 9 12 Pulser 11 13 Thermowell...

Страница 4: ... transit you should notify the shipper immediately file a claim with the carrier and notify your Elster supplier Damage to internal components may occur without visible external damage All new meters should be inspected and checked for free rotation of the impellers by lightly blowing into the inlet of the meter This slight air pressure should cause the impellers to rotate freely and come to a stop...

Страница 5: ...ex end of the meter 1 Ensure gas valves are closed and meter and piping are depressurized WARNING Failure to depressurize the meter prior to removing meter and or components could result in personal injury and or property damage 2 Remove oil fill plug in the counter end case cover using a 5mm hex key Figures 5 and 6 3 Using the supplied syringe and oil slowly add oil until it is 1 16 of the center ...

Страница 6: ...or 1 to 2 minutes to verify proper operation If the index does not start registration or if you hear knocking or scraping sounds coming from the meter stop the flow and follow appropriate decommissioning procedures before removing the meter from the line g If operation is satisfactory gradually open the meter s outlet valve to the full open position h Slowly close the bypass valve i Check the meter...

Страница 7: ... never left in a plug for a period of hours or days Severe deformation of the valves may occur if the probe is left in the plug for a long period of time b Differential Pressure Testing Typically when proving a meter both the meter pressure and meter differential pressure are used and recorded The RABO meter differential pressure ports are not full bore and most temperature probes will not fit thr...

Страница 8: ...0 40 50 60 75 100 150 175 250 290 0 0 0 1 0 3 0 7 1 4 2 1 2 8 3 4 4 1 5 2 6 9 10 3 12 1 17 2 20 0 100 110 130 160 230 300 370 430 500 600 770 1 110 1 270 1 780 2 050 3 500 3 900 4 600 5 800 8 200 10 500 12 900 15 300 17 700 21 200 27 200 39 100 45 000 62 800 72 300 5 500 6 100 7 200 9 100 12 800 16 600 20 300 24 000 27 800 33 400 42 700 61 400 70 700 98 700 113 700 160 170 200 260 360 470 570 680 ...

Страница 9: ... factor shown on the index face plate In the example on the next page the reading would be 1357900 cubic feet Figure 10 RABO Rotary Gas Meter 08 Elster Instromet c Dimensions and Weights b How to Read 11 Index ANSI Class 125 150 flat face flange connection A B C D E F G Nom Pipe Size Bolt Size H Bolts Flange Bolt Circle J Shipping Weight Carton Size in mm in mm in mm in mm in mm in mm in mm in mm ...

Страница 10: ...e change gear from the damaged index to the replacement Figures 14 and 15 Remove the change gear on the damaged index Pry the locking collar off with a knife blade or other sharp object Unscrew the change gear WARNING Each meter size uses a common index that has unique colored gears Interchanging colored gears will result in inaccurate readings Ensure the replacement index has the same color gears...

Страница 11: ... index cover 11 Install security seals LOCKING COLLAR CHANGE GEAR INDEX Meter Gear Color 3 5M G65 White 5 5M G100 Dark Green 9M G160 Green 14M G250 Red Table 1 LOCKING COLLAR CHANGE GEAR SUPPORT HOLE SUPPORT HOLE Figure 14 Change gear removal 12 Pulser All RABO meter indexes can be easily outfitted with a pulse output device to interface with auxiliary equipment Installation of a pulser is quick an...

Страница 12: ... Screw the thermowell into the meter and tighten to 18 ft lbs 8 Re pressurize the meter as instructed in Start up Commissioning see section 6 Figure 21 Auxilary equipment mounted ACCESSORY Figure 18 Auxiliary ports 13 Thermowell All RABO meters come equipped with two 2 NPT auxiliary ports on the meter body Figure 18 which can be used for sensing pressure and temperature of the flowing gas A thermo...

Страница 13: ...he posts in the index base plate 2 Ensure the shaft on the magnet holder aligns with the support hole on the index frame Figure 24 Figure 24 ID retaining screws INDEX RETAINING SCREW SUPPORT HOLE ON BACK SIDE INDEX RETAINING SCREW INDEX RETAINING SCREW 3 Tighten the three screws until snug Do not overtighten a Each screw should have a 2mm thick retaining washer to space the ID gearing and index as...

Страница 14: ...ent should be pipe or wall mounted adjacent to the meter Elster provides customized brackets for commonly used equipment Figure 32 Please contact your local Elster representative for details Figure 26 Support bracket attachment 3 5M 5 5M Horizontal Flow 3 5M 5 5M Vertical Flow 2 Using a 5mm Allen wrench loosely install two M6 socket head cap screws through the support bracket into the meter case c...

Страница 15: ...ng Impellers out of time Retime impellers Impellers contacting body Rotate manually to verify impellers spin freely Low registration Upstream or bypass leak Check all valves for leakage Non registration Broken or binding index odometer Replace odometer Obstruction in meter Remove meter remove obstructions flush meter 15 Troubleshooting probe and leading threads to ensure no pipe dope enters the met...

Страница 16: ...ms and solutions for the safe control and delivery of natural gas across the globe Trusted Brands Elster American Meter Company LLC Elster Perfection Corporation Elster Instromet Elster Canadian Meter Company LLC Elster Gas Depot Elster Meter Services Elster American Meter Company LLC 2221 Industrial Road Nebraska City NE 68410 USA T 1 402 873 8200 F 1 402 873 7616 www elster com gas 2014 by Elste...

Отзывы: