background image

AND ADJUSTMENT PROCEDURES FOR HONEYWELL 

 

OR MP-953F PNEUMATIC OPERATORS USED ON 

 HUMIDIFIERS

II.

INSTALLING THE PNEUMATIC OPERATOR

After necessary repairs have been made, perform the

following steps to install the operator.

1.

2.

3.

4.

5.

6.
7.

a.

10.

Press down the valve stem (16) so
that the valve is

completely

closed.

 Adapter

If the stem nut (17) and stem

 Nut

adapter (17A) were not removed 

 ---Stem

during

repairs,

check

the

adjustment shown

in Figure 2.

Adjust as required.

 surface

If the stem nut (17) and stem

for 

Operator

adapter (17A) were removed during

-Humidifier

repairs,

turn them onto the stem

and adjust as shown in Figure 2.

*With valve in closed position

Assemble the yoke assembly (36) and

Figure 2.

stem adapter (17A) in the following

manner.

Hold the bearing (36B) in the orientation shown in
Figure 1 (flat side down, depression up).

Slide

the bearing into the groove of the stem adapter

Place the spring (36C) on top of the stem adapter

 and bearing

(36B) with the longer tabs

pointed downward toward the notches in the sides

of the bearing.

c)

Slide the yoke (36A) over the spring 

bearing 

and stem adapter (17A) until the

short tabs of the spring snap into place on both
sides of the yoke.

Center the yoke over the stem

(16) 

l

Place the operator over the yoke assembly (36) and

lower the operator onto the bonnet (15).

Hold the operator to prevent it from turning.

Insert a

screwdriver into the slot in the top of the yoke (36)
and turn counterclockwise until the operator base (26)

contacts the mating surface of the bonnet (15).
Tighten 

 two set screws (27).

Load the stem by turning the yoke (36) clockwise until
the spring cup (28) is from 

 to 

 inch above the

operator base (26).

(Dimension

 in Figure 1).

Install the operator plug (35) in the top of the 

9530 or close the access plate on the top of the 
953B or MP-953F.

Turn on the steam to the humidifier.
Connect the air line to the operator and turn on the

instrument air.

Make any necessary adjustments.

2

Содержание MP953B

Страница 1: ...the operator from the bonnet installing the operator on the bonnet replacing steam seals changing operator springs replacing the diaphragm replacing the rolling seal and replacing the valve and seat I REMOVING THE PNEUMATIC OPERATOR Many repair procedures require that the operator be removed from the bonnet assembly To remove the operator perform the following steps Refer to Figure 1 see back cove...

Страница 2: ...36 36 34 3g 23 37 Ad y p L _ 33 I 14 ...

Страница 3: ... groove of the stem adapter 17A b Place the spring 36C on top of the stem adapter 17A and bearing 36B with the longer tabs pointed downward toward the notches in the sides of the bearing c Slide the yoke 36A over the spring 36C bearing 36B and stem adapter 17A until the short tabs of the spring snap into place on both sides of the yoke Center the yoke over the stem 16 l Place the operator over the...

Страница 4: ...th spring showing should be toward the valve Carefully guide the seals into place in the bonnet 15 Slide the packing gland 18 over the stem and turn it into the bonnet Tighten the packing gland securely Turn the bonnet assembly 15 into the humidifier cap and tighten securely Install the pneumatic operator in accordance with Section II IV CHANGING OPERATOR SPRlNGS 1 2 3 4 5 6 7 a 9 10 Remove the pn...

Страница 5: ... 8 Remove the pneumatic operator in accordance with Section I Perform Section IV CHANGING OPERATOR SPRINGS steps 2 through 8 to disassemble the operator With the operator in the upright position remove the three screws 23 and ring 39 that secure the rolling seal 22 Remove the rolling seal 22 Then separate the cup 37 from the old seal Place the cup 37 into the new rolling seal Install the new seal ...

Страница 6: ...Generally the diaphragm is replaced without removing the operator from the humidirier However there are times that the location of the humidifier makes it more convenient to remove the operator to a workbench for repair Remove the six cover screws 31 from the operator cover 33 Lift the cover 33 from the frame 29 Remove the hex nut 34 and washer 32 from the support 24 Do not disturb the yoke assemb...

Страница 7: ... Tighten the cover locking screw until it engages the relay housing Do not overtighten OPERATING 1 Set 2 S c TART POINT iDJUSTMENT COVER RANGE ADJUSTMENT r nn LOCKING hlY D SC R E W SCREWS RANGE ADJUSTMENT BLANK 3 SCREWS RELA BODY Fig 3 Adjustment Points of Gradutrol Relay START 3R RANGE AND START POINT MP 953F PILOT POSITIONER the start point on the positioner to the value desired Each click of t...

Страница 8: ... Replace the bias spring located on top of the operator e Replace the pilot positioner f Turn on the instrument air ADJUSTMENT CHECK MP 953B AND MP 953F MODELS ONLY 1 Install guages in the main and pilot air lines 2 Main air pressure should be equal to or more than top of sequencing range 13 psi is minimum main pressure for device to function 3 Slowly apply pilot pressure and note the pressure at ...

Отзывы: