IQ Force™ Dock Reference Manual
28
Dock Controls and Instrument Status
3.1.10 Gas Configuration
The calibration gas configuration for the dock is located on the right side
of the window. If the values are shown in red text, they can be changed.
If the values are shown in black text and you wish to change them, press
Change Config
.
NOTE
The gas values shown in the gas configuration screen
are the values that will be used in calibrating the IQ Force
instruments that interface with the dock regardless of the
IQ Force’s own built-in calibration gas settings.
To change the calibration-gas value, use the arrows to the right of the gas
value input box to increase or decrease the value.
To display LEL concentration in %CH
4
, check the
Display LEL as %CH4
.
To display in LEL, uncheck the box.
WARNING
Calibration values shown in the Gas Value column must match
those appearing on the calibra tion-gas cylinder(s) that will be
used to calibrate the detector. Non-matching calibration gas
and calibration-gas value settings will lead to inaccurate and
potentially dangerous readings.