IQ Force™ Dock Reference Manual
21
Initial PC–Dock Connection
Then this screen is displayed, indicating how much memory in the dock is
free for storing calibration records:
----------------
IQ Force Dock
100% Memory Free
----------------
The configuration of the gas connected to the dock is then displayed
(see
Section 3.1.10
), followed by the interval at which instruments will be
calibrated in the dock (see
Section 3.1.9
):
----------------
GAS CO/H2S/LEL
100/25/50
----------------
----------------
Dock Calibration
Interval=30 days
----------------
If the USB cable is not connected to the dock, the following three screens
are displayed:
----------------
Checking
Network Module
----------------
----------------
Configuring
Network Module
----------------
----------------
IP Address Is
151.150.10.201
----------------
When the dock is ready to accept an instrument, it alternates between the
following two screens:
----------------
READY 10:25*
20 APR 2012
----------------
----------------
READY 10:25*
CO/H2S/LEL
----------------
NOTE
The asterisk to the right of the time indicates that the
dock is connected to the PC and that the IQ Force Dock PC
software is running.
If the dock is running in standalone mode and has test records stored, a third
screen in the sequence will display the number of records.
----------------
READY 10:25
2 Records
----------------
Содержание IQ Force
Страница 1: ...IQ Force Dock Reference Manual ...
Страница 2: ......
Страница 54: ...IQ Force Dock Reference Manual 46 Help Menu ...
Страница 57: ...49 Calibration Frequency Recommendation IQ Force Dock Reference Manual Appendix ...
Страница 58: ......
Страница 59: ......
Страница 60: ...50109191 140 Revision 1 August 2012 2012 Honeywell Analytics All Rights reserved ...