46
www.tanksystem.com
10.11 Ullage/Interface verification
The sensitivity of the instrument in ullage / interface cannot be
adjusted. Both ullage and interface levels are set at the factory.
Checking ullage and interface level detection
The test liquid should be the one to be gauged. Fill in a
container with appropriate liquid.
Switch on the unit. The buzzer shall beep every 2 sec.
If the liquid is conductive (alcohol, water, ...)
• Check the
ullage
by immersing the ultrasonic gap sensor
but not the electrodes (position A); The buzzer shall beep
continuously.
• Check the
interface
by immersing the interface electrodes
(position B). The buzzer shall beep intermittently.
If the liquid is non conductive (gasoline, oil, ...)
• Check the
ullage
by immersing the sensor (position B); The
buzzer shall beep continuously.
• Check the interface by immersing the sensor (position B) in
water. The buzzer shall beep intermittently.
position A
position B
Figure 10-13
Содержание HERMetic UTImeter Gtex Chem
Страница 59: ... 59 TS 12100 TS 20620 TS 20618 TS 20538 11255 Issue 2 Figure 13 6 storage tube SS1 Q2 TS 10187 details ...
Страница 63: ... 63 ...
Страница 64: ...64 www tanksystem com ...
Страница 65: ... 65 ...
Страница 66: ...66 www tanksystem com ...
Страница 67: ... 67 ...
Страница 68: ...68 www tanksystem com ...
Страница 69: ... 69 ...
Страница 70: ...70 www tanksystem com ...
Страница 71: ... 71 ...
Страница 72: ...72 www tanksystem com ...
Страница 73: ... 73 ...
Страница 74: ...74 www tanksystem com ...
Страница 75: ... 75 15 Declaration of conformity ...