40
www.tanksystem.com
10.4 Tape replacement
THE REPLACEMENT OF THE TAPE DOES NOT REQUIRE
TO RE-CALIBRATE THE TEMPERATURE
.
Follow the different sequences as described below. The
Figure 13-1: general assembly, list of the main spare parts
can also help.
10.4.1 Disconnecting the tape from the sensor
Follow the instructions of section 10.5 “Sensing probe
replacement”.
10.4.2 Disconnecting the tape from the electronic box
• Unscrew with the 2.5 Allen key the 2 screws (A) of the battery
holder and pull it out as shown on Figure 10-5.
• Unscrew with the 2.5 Allen key the 4 screws (B) of the display
unit and pull it gently out as shown on Figure 10-5.
• Disconnect the connecting plug (C) as shown on Figure 10-6
and remove the display unit.
• Unscrew with the 2.5 Allen key the tape holder (G) by
removing the 2 screws (F) and the grounding cable (D) as
shown on Figure 10-6. Do not loose the 2 remaining screws
that secure the reel axle.
(A)
(B)
Figure 10-5
(C) connecting plug
to display unit
(D) grounding cable of tape
(E) black oversleeve of tape
(F) securing screw
(G) tape holder
(H) tape
(I) securing screw
Figure 10-6
Содержание HERMetic UTImeter Gtex Chem
Страница 59: ... 59 TS 12100 TS 20620 TS 20618 TS 20538 11255 Issue 2 Figure 13 6 storage tube SS1 Q2 TS 10187 details ...
Страница 63: ... 63 ...
Страница 64: ...64 www tanksystem com ...
Страница 65: ... 65 ...
Страница 66: ...66 www tanksystem com ...
Страница 67: ... 67 ...
Страница 68: ...68 www tanksystem com ...
Страница 69: ... 69 ...
Страница 70: ...70 www tanksystem com ...
Страница 71: ... 71 ...
Страница 72: ...72 www tanksystem com ...
Страница 73: ... 73 ...
Страница 74: ...74 www tanksystem com ...
Страница 75: ... 75 15 Declaration of conformity ...