Verification and adjustment of the reading
index
To verify or to adjust the reading index, in particular after having
renewed a tape, apply the following instruction:
• if the equipment is fitted with a 2” connector (Q2) remove
the clip and the collar as shown on Figure 10-12;
• unlock and remove the lid of the tape housing;
• put the tape cleaner on “DOWN” position;
• keep the equipment standing vertically on a flat surface;
• gently lower the tape until the sensor touches the surface
(Figure 10-12);
• adjust the index to the value corresponding to the connector
Q1 or Q2, as shown on Figure 10-12;
• In case of a 2” connector (Q2) put back the clip and the
collar.
• Put the housing lid back and lock it.
IMPORTANT NOTE
: these adjusting values for the reading
index are different from the heights shown in the section 8
“Examples of installation of the gauging system”. They take
into account the recessment of the reaction point from the
sensor tip end and other mechanical parameters.
Figure 10-12
SS1 Q1
SS1 Q2
SS2 Q2
Reading:
456 mm
(~1ft 5” 15/16)
Reading:
482 mm
(~1ft 7”)
Содержание HERMetic UTImeter Gtex Chem
Страница 59: ... 59 TS 12100 TS 20620 TS 20618 TS 20538 11255 Issue 2 Figure 13 6 storage tube SS1 Q2 TS 10187 details ...
Страница 63: ... 63 ...
Страница 64: ...64 www tanksystem com ...
Страница 65: ... 65 ...
Страница 66: ...66 www tanksystem com ...
Страница 67: ... 67 ...
Страница 68: ...68 www tanksystem com ...
Страница 69: ... 69 ...
Страница 70: ...70 www tanksystem com ...
Страница 71: ... 71 ...
Страница 72: ...72 www tanksystem com ...
Страница 73: ... 73 ...
Страница 74: ...74 www tanksystem com ...
Страница 75: ... 75 15 Declaration of conformity ...