Troubleshooting
Problem
Cause
Solution
"Battery 0%" message is
displayed
Depleted battery
Charge the rechargeable
battery pack
Error 1006
Temperature sensor fail
Replace PCBA
Error 1007
Data flash fail
Replace PCBA
Error 1008
BLE fail
Replace PCBA
Error 3001
RTC fail
Replace PCBA
Error 4004
The sensor is in the wrong
slot.
Correct the sensor position.
Error 4006
Sensors fail or no
communication
Replace the sensor or the
PCBA
Need Force Bump. "Bump
Now" message is displayed.
Bump overdue and must
carry out bump testing
before use.
Hold the button for 3 seconds
or connect to DC/SSDC or
insert to IntelliDoX to start the
bump testing; otherwise, the
detector will auto power off
after 60 secs.
Need Force Calibration
"Cal Now" message is
displayed.
Calibration overdue and
must carry out calibration
testing before use.
Hold the button for 3 seconds
or connect to DC/SSDC or
insert to IntelliDoX to start the
Calibration; otherwise, the
detector will auto power off
after 60 secs.
Detector alarms after start-
up sequence
Sensor not stabilized
O
2
sensor: Wait for at least 10
min before power on.
Sensors require calibration
For the NDIR-CH
4
sensor,
please allow 5 minutes
following startup for the
sensor to warm-up before
attempting to calibrate
Detector does not respond
when button is pressed
The battery state is critically
low, or the battery is
depleted.
Charge the rechargeable
battery pack
Detector does not respond
when button is pressed
Detector is performing
operations that do not
require user input.
Button operation restores
automatically when the
operation ends.
BW Flex Series
44
User Manual