Calibrate the Detector via DC App
1. Turn On the BW Flex Series detector and wait a few minutes to sensors warm up.
2. In your mobile device, open the Device Configurator app and pair with the detector. For
pairing details go to See "Bluetooth Pairing" on page
3. In your mobile device, tap on the Menu button and then select Remote Calibration
4. Tap Calibration, the IntelliFlash flashes amber to indicate the calibration process has started.
Note: If sensors fail to detect the gas, or cannot be spanned, repeat the calibration. If the
problem persists, replace the sensor.
5. Enter the Operator Name, and then Tap START. The detector will start Zero Calibration; the
IntelliFlash flashes amber.