Trouble shooting
Page 48
6.4 SPU status request (item QS)
The servo status request (item
QS
) consists of four bytes. Bytes 0, 1 and 2 are bit coded with information
about the level alarms, operational mode and general status. Byte 3 is an ASCII character that indicates the
active operational command.
Status byte 0:
Status byte 1:
bit 0 : low level alarm
bit 0 : displacer movement down
1 : low low level alarm
1 : displacer movement up
2 : high level alarm
2 : on level
3 : high high level alarm
3 : test flag
4 : motor limit switch low
4 : calibration test successful
5 : motor limit switch high
5 : calibration test failed
6 : 1
6 : 1
7 : 0
7 : 0
Status byte 2:
Active interface:
bit 0 : Active interface bit 0
bit 1 bit 0 Mode
1 : Active interface bit 1
0
0
I1
2 : dipped level
0
1
I2
3 : 0
1
0
I3
4 : general fail indication
1
1
dip
5 : no previous
ST
,
WD
or
SD
command
6 : 1
7 : 0
Status byte 3:
- :
no command active
N : interface profile
A :
balance test
L : lock test
B :
Block
M : measure frequency
C :
calibrate
R : tank profile
D :
go down
T : test gauge
F :
freeze
U : go up
Содержание 854 ATG Series
Страница 1: ...Instruction manual Series 854 ATG level gauge ...
Страница 2: ......
Страница 53: ...Appendix Instruction manual 854 ATG Page 51 ...
Страница 55: ...Appendix Instruction manual 854 ATG Page 53 ...
Страница 56: ...Appendix Page 54 Appendix C ASCII table ...
Страница 57: ...Appendix Instruction manual 854 ATG Page 55 Appendix D Display formats Level and Temperature display format A ...
Страница 58: ...Appendix Page 56 Level display format B Average gas temperature display format C ...
Страница 60: ...Appendix Page 58 HIMS pressure displays format F and H Servo density display format I ...
Страница 61: ...Appendix Instruction manual 854 ATG Page 59 Analog level output display format J Water level display format K ...
Страница 62: ...Appendix Page 60 Appendix E PCB layout XPU 2 HCU HCU SPU II GFC ...
Страница 67: ......