Trouble shooting
Page 46
6
Trouble shooting
The 854 ATG is an instrument with self diagnostics. Detected errors will be shown as status information on
the display (refer to Appendix D), or can be requested as items by the PET.
The following items contain the error codes of the processor boards:
EP
Error XPU(-2) request (communication processor unit)
ES
Error SPU request (servo processor unit)
These items contain an error code of the last error condition. The error codes can be read as long as the
gauge is not reset.
Besides the error codes, data items from level and optional functions (such as temperature and analog level
output) contain one or more status bytes which also give valuable information. These bytes are readable
ASCII characters. However, most of them are bit coded. Appendix C contains an ASCII table for conversion
of the status bits into the actual status.
An example for a bit coded status byte:
one (of the) status byte(s) reads:
F
;
written out in bits (refer to Appendix C):
0100 0110
;
(b7=0, b6=1, b5=0, b4=0, b3=0, b2=1, b1=1, b0=0).
Bit 7 is always a ‘0’ and bit 6 is always a ‘1’ to avoid ‘control’ characters.
Look up the relevant status byte in this section (e.g.
QS
in section 6.4) to determine the meaning of the bits
which are set to ‘1’. Only the bits set to ‘1’ represent an actual status.
6.1 Problems with displacer movement
If the displacer is not running freely, for instance stuck against a stilling well, it can be controlled manually.
Note:
Be aware that measuring wire is unrolled from the measuring drum. When the measuring wire is not kept
at tension, the result will be an uncontrolled wire movement which results in the worst case in the lost of
the measuring wire.
Item
Name
Description
W2
Protection level 2
Enter protection level 2 password.
SM
Set maintenance
Go into maintenance mode.
GD
Go down
Go down for approximately 200 mm (8").
FR
Freeze
Give the freeze command.
SO
Set operational
Exit maintenance mode
EX
Exit
Exit protection level 2
When hereafter the displacer is free and the weight is too high, the force transducer must be recalibrated.
When the measured displacer weight is too far out of range, the gauge is probably not levelled within 2
°
.
Then, improve the stability of the construction on which the gauge is mounted.
Содержание 854 ATG Series
Страница 1: ...Instruction manual Series 854 ATG level gauge ...
Страница 2: ......
Страница 53: ...Appendix Instruction manual 854 ATG Page 51 ...
Страница 55: ...Appendix Instruction manual 854 ATG Page 53 ...
Страница 56: ...Appendix Page 54 Appendix C ASCII table ...
Страница 57: ...Appendix Instruction manual 854 ATG Page 55 Appendix D Display formats Level and Temperature display format A ...
Страница 58: ...Appendix Page 56 Level display format B Average gas temperature display format C ...
Страница 60: ...Appendix Page 58 HIMS pressure displays format F and H Servo density display format I ...
Страница 61: ...Appendix Instruction manual 854 ATG Page 59 Analog level output display format J Water level display format K ...
Страница 62: ...Appendix Page 60 Appendix E PCB layout XPU 2 HCU HCU SPU II GFC ...
Страница 67: ......