11
Display Explanation Symbols:
If
and any of the following letters and numbers appear in the area
that systolic pressure should be displayed, an error has occurred with your
reading. See Troubleshooting section of this manual for more information.
User 1:
Appears when the monitor is operated by User 1.
User 2:
Appears when the monitor is operated by User 2.
Low Battery Symbol:
Appears when batteries should be replaced.
Pulse Symbol:
Shows the heart rate per minute.
Risk Category Index:
See page 29 for more information.
Irregular Heartbeat Detector:
See page 30 for more information.
Bluetooth
®
Symbol:
Appears when data is transmitting to your
mobile device.
•
Rest Assure ON:
The REST symbol will appear on monitor, then will countdown
for 5 minutes before starting single measurement.
•
Triple-Check Function
ON:
The REST symbol will flash among 3
measurements during countdown.
Memory Average:
Displays average of last 3 readings.
AM/PM Averaging:
Indicates the reading being displayed is an average from
the last 3 morning or last 3 evening measurements.
Multi-Read Symbol:
Appears when Triple-Check (Multi-Read) Function is
turned ON.
Average Symbol:
Displayed when viewing a Triple-Check average.
Measurement Indicator:
Indicates which measurement is being taken, or
which measurement is being viewed from a Triple-Check average reading.
Measurement Error:
Wrap the cuff correctly and keep arm steady during
measurement. Take new measurement.
Air Circuit Abnormality:
Check cuff connection. Take new measurement.
Pressure Exceeding 300 mmHg:
Turn the unit off to clear, then
take new measurement.
Error Determining Measurement Data:
Take new measurement.
Data Transmission Error:
Monitor cannot connect to the mobile device to
transmit data. Make sure
Bluetooth
is ON.
System Error:
Contact Consumer Relations.
/