WARNINGS and CAUTIONS
General
•
Failure to thoroughly review and adhere to the
information contained in this User Manual may pose a
potential hazard to the patient and/or the user and may
void the warranty.
•
The
TruNode
®
Gamma Probe System is not intended for
use in the central nervous system.
•
Do not modify this equipment without authorization of
the manufacturer
User Feedback Unit
•
Use only the Faxitron Power Supply provided with the
UFU for recharging. Use of another power supply may
increase the risk of electric shock
•
Fully charge the UFU before using the system.
•
Do not use the Charger in the operating room.
•
Do not use the Charger in the presence of an oxygen-
enriched environment or in the presence of a
flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
•
Keep the UFU off when charging.
•
Remove the Charger when the UFU is fully charged.
•
The UFU and Charger are non-sterile. Do not sterilize
these components.
•
If liquid is spilled on the UFU, remove it from service and
contact Faxitron for instructions to return for inspection.
Ph: 1. 877.910.0030.
•
Securely mount the UFU to an IV pole.