InSight FD Mini C-arm User Guide
Configuring the Application
36
MAN-04858 Revision 006
Table 19 Physician Preferences, Image Acquisition Tab, Section/Field Descriptions
Section/Field
Description
Snapshot Fluoro
Enables/disables Snapshot image acquisition mode.
Continuous
Fluoro
Enables/disables Continuous image acquisition mode.
Continuous
mode instant off
When checked and acquiring images in Continuous mode, X-rays
remain on until X-ray button or pedal is released. When
unchecked, tapping the X-ray button or pedal causes image
acquisition to behave like Snapshot mode; holding the button or
pedal pressed cause X-rays to remain on until released.
Default Image
Mode
Selects initial image mode setting for Image Acquisition screen.
X-ray Mode
Selects X-ray power level. Check Normal, IQ, or both. You must
check at least one mode or accept the system default. If you
uncheck both modes, the system enables Normal mode. The
system default enables Normal and IQ modes.
X-ray Technique
Slider control with 15 selectable settings from “Darker” to
“Brighter” for preferred selection of the kVp and mA under
automatic exposure control. Changing this value to “Darker”/
”Brighter” will marginally decrease/increase the kVp and mA
chosen by the automatic exposure control and, as a consequence,
the corresponding dose to the patient.
Simultaneous
kV/mA
When enabled, kV / mA goes up/down together to the next dose
curve pair when manually adjusting X-ray power. When disabled,
kV and mA can be adjusted independently.
Alarm on
Footswitch
Enables or disables setting of the alarm to sound when the X-ray
footswitch is depressed.
Time to X-ray
Alarm
Sets when the alarm will sound based on the accumulated X-ray
exposure time. Maximum value is 5 minutes.
Right Side
Display
Selects initial filmstrip/tools area setting for Image Acquisition
screen.
Thumbnail Hold
Determines the time a thumbnail image is displayed after the
acquisition ends.
Содержание Fluoroscan InSight FD Mini C-arm
Страница 1: ...InSight FD Mini C arm User Guide MAN 04858 Revision 006...
Страница 2: ......
Страница 3: ...InSight FD Mini C arm User Guide Part Number MAN 04858 Revision 006 July 2021...
Страница 34: ...InSight FD Mini C arm User Guide Configuring the Application 24 MAN 04858 Revision 006 Figure 12 System...
Страница 86: ...InSight FD Mini C arm User Guide System Maintenance 76 MAN 04858 Revision 006...
Страница 87: ......