![Hologic Fluoroscan InSight FD Mini C-arm Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/hologic/fluoroscan-insight-fd-mini-c-arm/fluoroscan-insight-fd-mini-c-arm_user-manual_2146527061.webp)
InSight FD Mini C-arm User Guide
Using the InSight FD Application
MAN-04858 Revision 006
51
Figure 29 Image Acquisition Screen Buttons
Table 32 Image Acquisition Screen, Button Functions
Button
Function
Acquisition Mode
Selection
Cycles through:
Continuous
– Displays continuous images when X-rays are
actuated.
Snapshot
– Captures and displays one image when X-rays are
actuated.
Cine Record
– Gathers all images acquired into a cine loop.
Continuous and/or Snapshot mode may be disabled (
NS (Noise
Suppression)
Cycles through the levels selected during configuration. See
Table 17 Physician Preferences, Section/Field Descriptions on Page
Select Ultra to acquire images for very still anatomy. Image
processing for Ultra level imaging averages frames at High noise
suppression.
Tag as Ref
Marks the image as a reference image for the current session or
unmarks a marked image.
Up
Increases the kV and/or mA incrementally by one.
X-ray Power
Adjustment
Selection
Cycles through:
Auto kV / mA
- Automatically adjusts kV / mA.
Manual kV / mA
(simultaneous kV/mA enabled) - Allow the user
to manually increase or decrease kV / mA.
Manual kV
(simultaneous kV/mA disabled) - Allow the user to
manually increase or decrease the kV.
Manual mA
(simultaneous kV/mA disabled) - Allow the user to
manually increase or decrease the mA.
Auto IQ kV / mA
- Automatically adjusts kV / mA. Select IQ mode
to acquire an enhanced image with improved clarity. Using IQ
mode increases the X-ray dose to the patient. Use Auto IQ when
enhanced clarity is required in accordance with ALARA.
on Page 33 for setting simultaneous
kV/mA.
Down
Decreases the kV and/or mA incrementally by one.
Reset Timer
Resets the X-ray warning timer to zero and turns off the audible
alarm.
Содержание Fluoroscan InSight FD Mini C-arm
Страница 1: ...InSight FD Mini C arm User Guide MAN 04858 Revision 006...
Страница 2: ......
Страница 3: ...InSight FD Mini C arm User Guide Part Number MAN 04858 Revision 006 July 2021...
Страница 34: ...InSight FD Mini C arm User Guide Configuring the Application 24 MAN 04858 Revision 006 Figure 12 System...
Страница 86: ...InSight FD Mini C arm User Guide System Maintenance 76 MAN 04858 Revision 006...
Страница 87: ......