background image

3

 

 

 

 

 

2.1 

PRECAUTIONS

 

2.1.1 

General

 

2.1.1.1 

Electrical shock hazard. Do not remove cover. Refer to authorized 
personnel for service.

 

2.1.1.2 

Surgitron

®

 Dual/EMC may present a hazard to patients with 

pacemakers. Consult qualified medical personnel.

 

2.1.1.3 

The manufacturer-supplied accessories should be used to ensure the 
proper operation of safety monitoring circuit.

 

2.1.1.4 

If higher than normal power settings are required, examine all cables, 
handpieces and the neutral plate for damage and confirm that cables are 
not wrapped around metal objects or touching the floor. In the event that 
higher power is still required, replace the neutral plate and handpieces.

 

2.1.1.5 

WARNING: No modification of this equipment is 
allowed.

 

2.1.1.6 

The patient should not come into contact with metal parts which are 
earthed or which have appreciable capacitance to earth. The use of 
antistatic sheeting is recommended for this purpose.

 

2.1.1.7 

Physician to patient skin to skin contact should be avoided. Use of 
insulating gloves is recommended. Patient skin-to-skin contact (eg, 
between the arms and body of the patient) should be avoided (eg, by 
insertion of dry gauze).

 

2.1.1.8 

When used simultaneously with physiological monitoring equipment, the 
monitoring electrodes on the patient should be placed as far as possible from 
the surgical electrodes. Do not use metal needle monitoring electrodes or 
EKG clip electrodes as burns may occur. Monitoring systems incorporating 
high-frequency current-limiting devices are recommended.

 

2.1.1.9 

The cables to the surgical electrodes should be positioned in such a way 
that contact with the patient or other leads is avoided. Temporarily 
unused active electrodes should be stored segregated from the patient.

 

2.1.1.10  The output power selected should be as low as possible for the 

intended purpose.

 

2.1.1.11  Apparent low output or failure of the surgical equipment to function 

correctly at the normal operating settings may indicate faulty 
application of the neutral plate or poor contact in its connection.

 

2.1.1.12  Before each use inspect the accessories, electrode cables 

and probes for possible physical damage.

 

2.1.1.13  Do not obstruct air circulation vents on the rear or underside of 

the generator.

 

2.1.1.14  Surgical gloves should be worn to prevent the User from 

becoming a part of the RF energy return pathway.

 

2.1.1.15  Use caution when treating subjects with metal implants. Metal implants in 

the electrical pathway (between the neutral plate and electrode) have the 
capacity to conduct energy and/or become warm.

 

2.1.1.16  High frequency surgical equipment output from either active 

electrode may change when operating mode is changed.

 

2.1.1.17  Select associated equipment and active accessories that have rated 

accessory voltages equal to or greater than the maximum output voltage. 
Avoid high frequency voltage settings where maximum output voltage may 
exceed rated accessory voltage.

 

2.1.1.18  For surgical procedures where the high frequency current could flow 

through parts of the body having a relatively small cross-sectional area, 
the use of bipolar techniques may be desirable to avoid unwanted 
coagulation.

 

2.1.1.19  Similar to all energy sources, by-products such as tissue smoke plume 

and aerosols, create a health concern about their carcinogenic and 
infectious potential. Appropriate protective measures including effective 
smoke evacuation should always be utilized.

 

 

 

Содержание CYNOSURE ELLMAN SURGITRON 4.0 Dual RF/90 IEC

Страница 1: ...4 0 Dual RF 90 IEC Instruction Manual English English...

Страница 2: ...subjected to an accident tampering alteration misuse abuse negligence fire or water damage Warranty does not cover defects or damage caused directly or indirectly as a result of service by unauthoriz...

Страница 3: ...FOR USE Page 12 5 0 Preparation for Use 5 1 Waveform Selection 5 2 Smoke Hazard VI SURGICAL DESCRIPTION AND USE Page 14 6 0 Learning to Use Radiosurgery 6 1 Pre Operative Cutting Practice 6 2 Pre Oper...

Страница 4: ...stics 8 1 Preventive Inspection and Maintenance 8 2 Safety Features 8 3 Input Characteristics 8 4 Size and Weight 8 5 Operating Conditions 8 6 Transportation Storage Conditions 8 7 Error Codes Table I...

Страница 5: ...arranty II SAFETY Improper use of this or any high frequency surgical instrument can result in a patient injury Read and become familiar with all directions before using this product 2 0 WARNINGS 2 0...

Страница 6: ...with the patient or other leads is avoided Temporarily unused active electrodes should be stored segregated from the patient 2 1 1 10 The output power selected should be as low as possible for the int...

Страница 7: ...as monitoring electrodes 2 1 3 Flammable Environment 2 1 3 1 Some materials for example cotton wool and gauze when saturated with oxygen may be ignited by sparks produced in normal use of the equipmen...

Страница 8: ...e of this device in the unsorted municipal waste stream 3 2 CLEANING DISINFECTING AND STERILIZING THE UNIT AND ACCESSORIES 3 2 1 Cleaning and disinfecting the unit Cleaning and disinfecting of the uni...

Страница 9: ...available for use with the Surgitron Dual EMC A three button handpiece adds convenience by enabling the user to control CUT BLEND and COAG modes via fingerswitch The depressed button will control out...

Страница 10: ...er cord supplied by the manufacturer Line Fuses The unit is protected with two line fuses If a fuse blows the unit should be checked for possible faults before the fuse is replaced and the unit is ope...

Страница 11: ...the top of the line fuse door Remove two fuse holders Replace fuses with the same value size fuses ONLY The fuse holder arrows should be pointing towards the right Snap the compartment back with the f...

Страница 12: ...If the neutral plate is not properly connected to the patient and or Surgitron Dual EMC the Neutral LED Indicator will turn red and the audio alarm will sound until the patient return neutral plate is...

Страница 13: ...tch Handpiece Receptacle accepts any monopolar handpiece that is compatible with the Surgitron Dual EMC Yellow Footswitch Indicator The light indicator will be ON only when the CUT control footswitch...

Страница 14: ...gures in Section 7 20 Power up Button for Coagulation Mode A single depression of the button switch increases the power setting by one Continuous depression gradually increases the power to maximum 10...

Страница 15: ...ent s body as close as possible to the operating field as per the Safety Instructions specified in Section 2 1 4 5 0 5 Select the desired electrode and handpiece and connect them as instructed by thei...

Страница 16: ...ated Waveform is an intermittent flow of the high frequency radiowaves that produces hemostasis The Partially Rectified Modulated Wave is also recommended for coagulating blood vessels via an indirect...

Страница 17: ...point where excessive dehydration occurs Preventing the accumulation of such heat is a basic objective of radiosurgical technique The spread of lateral heat in tissue depends upon various factors as i...

Страница 18: ...e to the electrode while cutting then increase the power gradually until a smooth cut occurs H If the setting is too high then you will notice sparking and discoloration along the electrode path Reduc...

Страница 19: ...ed wave with different techniques Bleeding should be stopped by some form of direct manually applied pressure When bleeding has momentarily stopped final sealing of the vessels can be accomplished by...

Страница 20: ...urgitron Dual EMC CUT MODE Figure 7 1 RESISTIVE LOAD OHMS POWER WATTS POWER WATTS Half Setting 50 Full Setting 100 Typical Nominal Plot CUT MODE Power Output vs Resistive Load Typical Nominal Plot 90w...

Страница 21: ...ODE Figure 7 3 RESISTIVE LOAD OHMS RESISTIVE LOAD OHMS POWER WATTS POWER WATTS Half Setting 50 Full Setting 100 Typical Nominal Plot COAG MODE Power Output vs Resistive Load Half Setting 50 Full Setti...

Страница 22: ...MODE Figure 7 5 RESISTIVE LOAD OHMS RESISTIVE LOAD OHMS POWER WATTS POWER WATTS Typical Nominal Plot BIPOLAR MODE Power Output vs Resistive Load Half Setting 50 Full Setting 100 Half Setting 50 Full S...

Страница 23: ...20 Surgitron Dual EMC MONOPOLAR BIPOLAR MODES Figure 7 6 Typical Nominal Plot Voltage Output vs Digital Setting Open Circuit No Load CUT CUT COAG HEMO FULG BIPOLAR VOLTAGE PEAK PEAK DIGITAL SETTING...

Страница 24: ...and method of circuit connection RF Power out is within 20 of nominal above when greater than 10 of maximum RF Power out is stable within 10 within 10 seconds after activation RF output frequency is w...

Страница 25: ...s above 85 C the unit will declare an error message error code 12 8 2 7 Continuous power test monitor output power Error code 13 if power test failed 8 2 8 Error indication after self check Refer to l...

Страница 26: ...had a warm reset 4 A D converter error 7 Foot Fingerswitch interface failed 8 Bond sensor interface failed 9 Keypad pressed during power up protection failed 10 Foot or Fingerswitch pressed during pow...

Страница 27: ...f interference plug the unit in a separate outlet or plug the unit in another room 9 1 2 Medical electrical equipment requires special precautions regarding electromagnetic compatibility EMC and needs...

Страница 28: ...overed with synthetic material the relative humidity should be kept at levels to reduce electrostatic charge to suitable levels electrical fast transients bursts IEC 61000 4 4 2 kV for power supply li...

Страница 29: ...ce in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in t...

Страница 30: ...ent Guidance 385 TETRA 400 380 390 MHz 27 V m 27 V m Unless otherwise specified the equipment was tested at a distance of 0 3m The equipment was tested using the IEC 61000 4 3 test method 450 GMRS 460...

Страница 31: ...12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transm...

Страница 32: ...Hologic BVBA Da Vincilaan 5 1930 Zaventem Belgium Tel 32 2 711 46 80 Cynosure LLC 5 Carlisle Road Westford MA USA Tel 888 523 2233 Fax 978 349 6152 DIM 50 65 Rev 002 04 2019...

Отзывы: