background image

10

 

 

 

 
 

Units are shipped from the factory with default power levels of 0 and modes of 
Cut and Hemo. To modify default values, simultaneously press and hold the 
Cutting Mode and Coagulation Mode Select buttons while turning on the unit. 
When CLR is displayed release the buttons and select the desired settings. Turn 
off power.

 

 

Fault LED Indicator:

 

The Fault LED Indicator will be Red if any error is detected; otherwise, it is Green.

 

Neutral Plate Receptacle:

 

The neutral plate must be applied to the patient and connected to the equipment 
for monopolar techniques. The Surgitron

®

 Dual/EMC is equipped with a neutral 

plate safety monitor which monitors the electrical connection between the neutral 
plate and the unit, as well as between the neutral plate and the patient.

 

 

Neutral LED Indicator:

 

The indicator lights up Red and triggers an audible alarm if the neutral plate is not 
properly applied. The LED indicator will turn to Green when the neutral plate is 
correctly applied. This function is automatically bypassed if Bipolar mode is 
selected.

 

 

Bipolar Cord Receptacle:

 

Bipolar function can be activated by footswitch only.

 

 

Fingerswitch/Handpiece Receptacle:

 

The Fingerswitch/Handpiece Receptacle accepts any monopolar handpiece that 
is compatible with the Surgitron

®

 Dual/EMC.

 

 

Yellow Footswitch Indicator: 

The light indicator will be ON only when the CUT 

control footswitch pedal is depressed in Surgitron

®

 Dual/EMC model.

 

 

Blue Footswitch Indicator:

 

The light indicator will be ON only when the COAG control footswitch pedal is 
depressed in Surgitron

®

 Dual/EMC model.

 

 

Cutting Mode Function Select Button: 

It is for selecting Cut or Blend functions.

 

Yellow Light (Cut) Indicator: 

The indicator illuminates when Cut function has 

been selected.

 

 

Yellow Light (Blend) Indicator: 

The indicator illuminates when Blend function 

has been selected.

 

 

Output Power Intensity Indicator for Cutting Mode: 

The default state is zero 

when On/Off power switch is turned on. The output power will remain at the 
selected level unless it is changed. The number displayed indicates the relative 
power, which will be delivered to the patient when the mode is activated. For true 
output power, see figures in Section 7.

 

 

Power-up Button for Cutting Mode: 

A single depression of the button switch 

increases the power setting by one. Continuous depression gradually increases 
the power to maximum - 100.

 

 

Power-down Button for Cutting Mode: 

A single depression lowers the power 

setting by one. Continuous depression gradually decreases the power to 
minimum - 0

.

 

 

 

2

 

4

 

5

 

6

 

7

 

8

 

10

 

11

 

12

 

13

 

14

 

Содержание CYNOSURE ELLMAN SURGITRON 4.0 Dual RF/90 IEC

Страница 1: ...4 0 Dual RF 90 IEC Instruction Manual English English...

Страница 2: ...subjected to an accident tampering alteration misuse abuse negligence fire or water damage Warranty does not cover defects or damage caused directly or indirectly as a result of service by unauthoriz...

Страница 3: ...FOR USE Page 12 5 0 Preparation for Use 5 1 Waveform Selection 5 2 Smoke Hazard VI SURGICAL DESCRIPTION AND USE Page 14 6 0 Learning to Use Radiosurgery 6 1 Pre Operative Cutting Practice 6 2 Pre Oper...

Страница 4: ...stics 8 1 Preventive Inspection and Maintenance 8 2 Safety Features 8 3 Input Characteristics 8 4 Size and Weight 8 5 Operating Conditions 8 6 Transportation Storage Conditions 8 7 Error Codes Table I...

Страница 5: ...arranty II SAFETY Improper use of this or any high frequency surgical instrument can result in a patient injury Read and become familiar with all directions before using this product 2 0 WARNINGS 2 0...

Страница 6: ...with the patient or other leads is avoided Temporarily unused active electrodes should be stored segregated from the patient 2 1 1 10 The output power selected should be as low as possible for the int...

Страница 7: ...as monitoring electrodes 2 1 3 Flammable Environment 2 1 3 1 Some materials for example cotton wool and gauze when saturated with oxygen may be ignited by sparks produced in normal use of the equipmen...

Страница 8: ...e of this device in the unsorted municipal waste stream 3 2 CLEANING DISINFECTING AND STERILIZING THE UNIT AND ACCESSORIES 3 2 1 Cleaning and disinfecting the unit Cleaning and disinfecting of the uni...

Страница 9: ...available for use with the Surgitron Dual EMC A three button handpiece adds convenience by enabling the user to control CUT BLEND and COAG modes via fingerswitch The depressed button will control out...

Страница 10: ...er cord supplied by the manufacturer Line Fuses The unit is protected with two line fuses If a fuse blows the unit should be checked for possible faults before the fuse is replaced and the unit is ope...

Страница 11: ...the top of the line fuse door Remove two fuse holders Replace fuses with the same value size fuses ONLY The fuse holder arrows should be pointing towards the right Snap the compartment back with the f...

Страница 12: ...If the neutral plate is not properly connected to the patient and or Surgitron Dual EMC the Neutral LED Indicator will turn red and the audio alarm will sound until the patient return neutral plate is...

Страница 13: ...tch Handpiece Receptacle accepts any monopolar handpiece that is compatible with the Surgitron Dual EMC Yellow Footswitch Indicator The light indicator will be ON only when the CUT control footswitch...

Страница 14: ...gures in Section 7 20 Power up Button for Coagulation Mode A single depression of the button switch increases the power setting by one Continuous depression gradually increases the power to maximum 10...

Страница 15: ...ent s body as close as possible to the operating field as per the Safety Instructions specified in Section 2 1 4 5 0 5 Select the desired electrode and handpiece and connect them as instructed by thei...

Страница 16: ...ated Waveform is an intermittent flow of the high frequency radiowaves that produces hemostasis The Partially Rectified Modulated Wave is also recommended for coagulating blood vessels via an indirect...

Страница 17: ...point where excessive dehydration occurs Preventing the accumulation of such heat is a basic objective of radiosurgical technique The spread of lateral heat in tissue depends upon various factors as i...

Страница 18: ...e to the electrode while cutting then increase the power gradually until a smooth cut occurs H If the setting is too high then you will notice sparking and discoloration along the electrode path Reduc...

Страница 19: ...ed wave with different techniques Bleeding should be stopped by some form of direct manually applied pressure When bleeding has momentarily stopped final sealing of the vessels can be accomplished by...

Страница 20: ...urgitron Dual EMC CUT MODE Figure 7 1 RESISTIVE LOAD OHMS POWER WATTS POWER WATTS Half Setting 50 Full Setting 100 Typical Nominal Plot CUT MODE Power Output vs Resistive Load Typical Nominal Plot 90w...

Страница 21: ...ODE Figure 7 3 RESISTIVE LOAD OHMS RESISTIVE LOAD OHMS POWER WATTS POWER WATTS Half Setting 50 Full Setting 100 Typical Nominal Plot COAG MODE Power Output vs Resistive Load Half Setting 50 Full Setti...

Страница 22: ...MODE Figure 7 5 RESISTIVE LOAD OHMS RESISTIVE LOAD OHMS POWER WATTS POWER WATTS Typical Nominal Plot BIPOLAR MODE Power Output vs Resistive Load Half Setting 50 Full Setting 100 Half Setting 50 Full S...

Страница 23: ...20 Surgitron Dual EMC MONOPOLAR BIPOLAR MODES Figure 7 6 Typical Nominal Plot Voltage Output vs Digital Setting Open Circuit No Load CUT CUT COAG HEMO FULG BIPOLAR VOLTAGE PEAK PEAK DIGITAL SETTING...

Страница 24: ...and method of circuit connection RF Power out is within 20 of nominal above when greater than 10 of maximum RF Power out is stable within 10 within 10 seconds after activation RF output frequency is w...

Страница 25: ...s above 85 C the unit will declare an error message error code 12 8 2 7 Continuous power test monitor output power Error code 13 if power test failed 8 2 8 Error indication after self check Refer to l...

Страница 26: ...had a warm reset 4 A D converter error 7 Foot Fingerswitch interface failed 8 Bond sensor interface failed 9 Keypad pressed during power up protection failed 10 Foot or Fingerswitch pressed during pow...

Страница 27: ...f interference plug the unit in a separate outlet or plug the unit in another room 9 1 2 Medical electrical equipment requires special precautions regarding electromagnetic compatibility EMC and needs...

Страница 28: ...overed with synthetic material the relative humidity should be kept at levels to reduce electrostatic charge to suitable levels electrical fast transients bursts IEC 61000 4 4 2 kV for power supply li...

Страница 29: ...ce in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in t...

Страница 30: ...ent Guidance 385 TETRA 400 380 390 MHz 27 V m 27 V m Unless otherwise specified the equipment was tested at a distance of 0 3m The equipment was tested using the IEC 61000 4 3 test method 450 GMRS 460...

Страница 31: ...12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transm...

Страница 32: ...Hologic BVBA Da Vincilaan 5 1930 Zaventem Belgium Tel 32 2 711 46 80 Cynosure LLC 5 Carlisle Road Westford MA USA Tel 888 523 2233 Fax 978 349 6152 DIM 50 65 Rev 002 04 2019...

Отзывы: