9
4.2
FRONT PANEL CONTROL ELEMENTS
Power Switch:
When the power switch is turned on, the unit automatically performs the following functions:
1 = ON
0 = OFF
•
The audio signal is sounded.
•
All start up sequences of tones on the front panel are activated during the
functional test.
•
All front panel LEDS are illuminated.
•
The patient return neutral plate is tested. If the neutral plate is not properly
connected to the patient and/or Surgitron
®
Dual/EMC, the Neutral LED Indicator
will turn red and the audio alarm will sound until the patient return neutral plate is
properly applied. The test is bypassed when the bipolar function is selected. All
power digital indicators shall be at the default state. If any error code is shown, the
error must be cleared prior to using the equipment. (For error code information,
see 8.7 - Error Codes Table).
Figure 4.2
Front Panel
FINGERSWITCH
HANDPIECE
BIPOLAR
HANDPIECE
NEUTRAL
PLATE
1